5
3. Installation
3.1 Battery
Handpiece has built-in lithium battery. It charges wirelessly through the base.
•
Plug in power adapter and connect it to the charging base
•
Seat handpiece on the charging base
•
Flashing yellow light indicates device charging.
•
When fully charged, yellow light will stay on.
3.2 Contra angle
•
The contra angle adopts precision gear transmission, and the transmission ratio is
1:1.
• Clean and disinfect with disinfectant of neutral PH value before the first use.
•
Clean and disinfect after usage on each patient
•
Sterilize under 134°C, 2.0~2.3bar (0.20~0.23MPa).
• Contra angle in this package only fits ApexPilot G1. This will not fit other
handpieces.
Installation of contra angle
•
Align the pin of the contra-
angle with one of the holes on
the motor handpiece, and push
in the contra-angle until there is
an audible click
•
A “click” sound indicates that
the installation is in place.
Note:
There are 6 holes on the motor
handpiece which allows different
orientations and rotational range
to set the contra angle at
Once chosen orientation is set,
orientation is fixed. Contra angle
cannot rotate.
Removal of contra angle
Pull out the contra angle horizontally when the motor handpiece is off.