TECHNICAL MANUAL
DGG 700 B
142
PATIENT
IMPEDANCE
Measurement of the
patient impedance has
not been correctly
carried out
Check the instructions and carry out
the Hardware test again in order to
calibrate the impedance meter
Check the continuity of the patient
electrodes connection cable
Check the patient electrodes
connector to see if it is damaged
and/or broken
Check the connection of the
electrodes connector to the Power
Board (APEX and EXT connectors)
Check or replace the CPU board
MONITOR
The signal acquired by
means of the
disposable electrodes
or the paddles is of bad
quality
Check and if necessary replace the
paddles or the disposable electrodes
cable
Check and if necessary replace the
multifunction connector (MFC)
Check the connection of the
multifunction connector to the CPU
board
Check or replace the CPU board
The signal acquired
using the patient cable
is of bad quality
Check and if necessary replace the
patient cable
Check and if necessary replace the
patient cable connector
Check the connection of the patient
cable connector to the CPU board
Check or replace the CPU board
The % SpO2 value is
not monitored
Check or replace the pulse oximetry
sensor and/or the spiral cable
Check and replace the patient cable
connector if necessary
Check the connection between the
connector and the SpO2 board
Check the connection between the
SpO2 board and the CPU board
Check or replace the SpO2 board
Check or replace the CPU board
Summary of Contents for Reanibex 700
Page 1: ......
Page 2: ......
Page 3: ...REANIBEX Serie 700 Technical Manual Approved R D Director Revision H Date December 2011...
Page 4: ......
Page 6: ...TECHNICAL MANUAL DGG 700 B...
Page 10: ...TECHNICAL MANUAL DGG 700 B iv Blank page...
Page 42: ...TECHNICAL MANUAL DGG 700 B 32 Blank page...
Page 51: ...TECHNICAL MANUAL DGG 700 B 41...
Page 54: ...TECHNICAL MANUAL DGG 700 B 44 Impedancia Paciente de 50 Impedancia Paciente de 75...
Page 55: ...TECHNICAL MANUAL DGG 700 B 45 Impedancia Paciente de 100 Impedancia Paciente de 125...
Page 56: ...TECHNICAL MANUAL DGG 700 B 46 Impedancia Paciente de 150 Impedancia Paciente de 175...
Page 61: ...TECHNICAL MANUAL DGG 700 B 51 Blank page...
Page 63: ...TECHNICAL MANUAL DGG 700 B 53 Blank page...
Page 103: ...TECHNICAL MANUAL DGG 700 B 93 Blank page...
Page 113: ...TECHNICAL MANUAL DGG 700 B 103 Blank page...
Page 121: ...TECHNICAL MANUAL DGG 700 B 111 Blank page...
Page 125: ...TECHNICAL MANUAL DGG 700 B 115 Blank page...
Page 156: ...TECHNICAL MANUAL DGG 700 B 146 Blank page...
Page 162: ...TECHNICAL MANUAL DGG 700 B 152 Blank page...
Page 164: ...TECHNICAL MANUAL DGG 700 B 154 Blank page...
Page 167: ...TECHNICAL MANUAL DGG 700 B 157 A 4 Assembly Diagrams...