TECHNICAL MANUAL
DGG 700 B
119
-
Discharged energy and biphasic pulse phases time -
It verifies during the
discharge, that the phase times of both the biphasic pulse and the discharged
energy are within the established limits.
9-
PATIENT MONITORING -
Is entrustd with testing the part of the equipment in
charge of monitoring the parameters of the patient.
-
COMMUNICATION with the ADuC -
The CPU tries series communication
with the ADuC. If there is no response within a established time, it considers
that it is not possible to establish communication and returns an error.
-
A/D and D/A CONVERTER -
In this test it is checked that the converters of
the equipment work correctly.
-
Patient IMPEDANCE meter circuit -
By means of this Test the impedance
measurement circuit is tested and calibrated.
-
ECG CIRCUIT Testing -
By means of this Test the impedance measurement
circuit is tested and calibrated
-
LOOSE ELECTRODE DETECTION CIRCUIT -
This test checks that the
loose electrode detection circuit works correctly, that is that it is capable of
detecting this condition
-
Communication with the paddles
- The purpose of this test is to check the
correct operation of the I2C bus that communicates with the paddles
10-
COPROCESSOR: Presence and communication
The CPU tries to establish series communication with the coprocessor
(COP_RXD0 and COP_TXD0). If it does not respond because it does not exist or
there is some communication problem it will give an error.
11-
PACEMAKER -
During this test apart from comunication with the Pacemaker, the
following points are tested:
-
125 V Source -
The control signal will be viewed to see if the source works or
not.
-
Pulse current -
The current delivered trough a knwn test impedance is
monitored to verify if it is the same to the one selected during the test.
-
Pulse time
- Checks that the duration of the pulse provided by the pacemaker is
correct and does not exceed the established limits.
Summary of Contents for Reanibex 700
Page 1: ......
Page 2: ......
Page 3: ...REANIBEX Serie 700 Technical Manual Approved R D Director Revision H Date December 2011...
Page 4: ......
Page 6: ...TECHNICAL MANUAL DGG 700 B...
Page 10: ...TECHNICAL MANUAL DGG 700 B iv Blank page...
Page 42: ...TECHNICAL MANUAL DGG 700 B 32 Blank page...
Page 51: ...TECHNICAL MANUAL DGG 700 B 41...
Page 54: ...TECHNICAL MANUAL DGG 700 B 44 Impedancia Paciente de 50 Impedancia Paciente de 75...
Page 55: ...TECHNICAL MANUAL DGG 700 B 45 Impedancia Paciente de 100 Impedancia Paciente de 125...
Page 56: ...TECHNICAL MANUAL DGG 700 B 46 Impedancia Paciente de 150 Impedancia Paciente de 175...
Page 61: ...TECHNICAL MANUAL DGG 700 B 51 Blank page...
Page 63: ...TECHNICAL MANUAL DGG 700 B 53 Blank page...
Page 103: ...TECHNICAL MANUAL DGG 700 B 93 Blank page...
Page 113: ...TECHNICAL MANUAL DGG 700 B 103 Blank page...
Page 121: ...TECHNICAL MANUAL DGG 700 B 111 Blank page...
Page 125: ...TECHNICAL MANUAL DGG 700 B 115 Blank page...
Page 156: ...TECHNICAL MANUAL DGG 700 B 146 Blank page...
Page 162: ...TECHNICAL MANUAL DGG 700 B 152 Blank page...
Page 164: ...TECHNICAL MANUAL DGG 700 B 154 Blank page...
Page 167: ...TECHNICAL MANUAL DGG 700 B 157 A 4 Assembly Diagrams...