3. Safety information and warnings

 Warning

•  Before use, ensure that there is no visible damage to 

the unit or accessories. When in doubt, do not use 

the unit and contact your dealer or the customer ser-

vice address provided.

• Using the unit is not a substitute for consultation with 

or treatment from a physician. Whenever you have 

any pain or illness, always contact your physician first.

•  If you have health concerns of any kind, contact your 

general practitioner!

• When using the nebuliser, always follow the basic 

hygiene procedures.

• Always follow your doctor’s instructions regarding 

the type of medicine to be used and the dose, fre-

quency, and duration of inhalation.

• Only use medicines that have been prescribed or rec-

ommended by your doctor or pharmacist.

• If the unit does not function correctly or if you begin to 

feel unwell or feel pain, stop using the unit immediately.

• During use, hold the unit away from the eyes, as some 

nebulised medicines may cause damage to the eyes.

•  Do not operate the unit in the presence of flammable 

gases.

•  Do not use the device near electromagnetic ransmit-

ters.

• This device is not intended to be used by children or 

persons with restricted physical, sensory (e.g. insen-

sitivity to pain) or mental abilities or persons lacking 

in the required experience or knowledge for safe op-

eration of the device, unless supervised or instructed 

by a person responsible for their safety.

• Keep children away from packaging materials (risk 

of suffocation).

• Do not use attachments that are not recommended 

by the manufacturer.

• Do not dip the unit into water and do not use it in wet 

rooms. Do not allow any liquid to penetrate the unit.

• Protect the unit from strong impacts.

• Do not use the unit if it has fallen or been dropped 

on the floor, been exposed to extreme humidity or 

is otherwise damaged. In case of doubt, contact our 

customer service department or your local dealer.

•  The IH 50 must only be operated using the appropri-

ate Beurer accessories.The use of other accessories 

can impair therapeutic efficiency and may damage 

the device.

• Remove the batteries from the device if you will not 

be using it for an extended period. Leaking batteries 

may damage the device.

 Important

•  Power cuts, sudden faults or other unfavourable con-

ditions may render the unit unusable. You are there-

fore recommended to keep a spare device or re-

placement medicine (as agreed with the doctor).

• The device must be stored away from sources of 

heat.

• Do not use the unit in a room in which a spray has 

previously been used. If this is the case, ventilate the 

room before commencing treatment.

• For reasons of hygiene, each user should use their 

own set of accessories (mesh atomiser, mask and 

mouthpiece).

• Store the device in a location protected against en-

vironmental influences. The device must be stored 

in the specified ambient conditions.

• Do not use any liquids that have a viscosity (thick-

ness) greater than 3, as these may damage the mesh 

irreparably.

• Do not use any medicines in powder form (even if 

dissolved).

• Do not shake the atomiser, as this may cause liquid 

to leak, and the device may become damaged.

Power pack

 Important

• Do not use any power pack other than the one sup-

plied.

•  Ensure that the mains socket cannot come into con-

tact with liquids.

• If adapters or extensions are required, these must 

comply with the applicable safety regulations. The 

electrical capacity and the maximum capacity speci-

fied on the adapter must not be exceeded.

•  After using the power pack, unplug it from the socket.

• Connect the unit only to the mains voltage listed on 

the type plate.

General notes

 Important

• Use the unit only:

 – On humans,

  –  for its intended purpose (aerosol inhalation) and in 

the manner described in these instructions for use.

• Any improper use can be dangerous!

• In acute emergencies, first aid takes priority.

• Apart from medicines, use only a saline solution.

• This unit is not intended for commercial or clinical 

use, but only for individual, private household use!

Summary of Contents for IH50

Page 1: ...er Instructions for use 12 F Inhalateur Mode d emploi 21 E Inhalador Manual de instrucciones 31 I Inalatore Istruzioni per l uso 41 T Inhalat r Kullan m k lavuzu 50 r 59 Q Inhalator Instrukcja obs ugi...

Page 2: ...d atmosphere Inhale slowly and deeply to enable the medicine to reach the small bronchi deep into the lungs Exhale normally Once prepared the device is suitable for reuse Recon ditioning includes the...

Page 3: ...our local dealer The IH 50 must only be operated using the appropri ate Beurer accessories The use of other accessories can impair therapeutic efficiency and may damage the device Remove the batteries...

Page 4: ...e protective gloves Repairs Note Never open or attempt to repair the unit yourself as otherwise proper function is no longer guaranteed Fail ure to observe this regulation shall void the warranty If y...

Page 5: ...ration on the bottom of the casing Fold the battery com partment lid down again and push it back in place in the opposite direction to the arrows Note When changing batteries ensure that the medicine...

Page 6: ...they can enter the air ways and lungs you must inhale slowly and deeply hold your breath briefly 5 to 10 seconds and then exhale quickly You should only use nebulisers to treat respiratory disorders...

Page 7: ...be poisonous when in contact with the skin or mucous membranes or when swallowed or inhaled If required you can clean the device casing with a lightly moistened cloth that has been soaked in mild soap...

Page 8: ...mpletely by shaking it Otherwise nebulisation may not function properly after reassembling the device If this is the case disassemble the mesh atomiser 3 again and shake it so that the water can escap...

Page 9: ...frequency 100 kHz Housing material ABS TPE Mains connection 100 240 V 50 60 Hz 0 15 A Operating conditions Temperature 5 C 40 C Relative humidity 80 non con densing Storage and transport conditions Te...

Page 10: ...ion of this device 11 Disposal Battery disposal The empty completely flat batteries must be dis posed of through specially designated collection boxes recycling points or electronics retailers You are...

Page 11: ...it s effectuer que sur avis m dical Soyez calme et d tendu pour proc der l inhalation Inspirez lentement et profond ment afin que le m dicament parvienne jusqu aux bronches les plus fines et les plus...

Page 12: ...ant N immergez pas l appareil sous l eau et ne l utilisez pas dans des locaux humides Aucun liquide ne doit p n trer dans l appareil Prot gez l appareil contre les chocs importants Si l appareil est t...

Page 13: ...g es ni court cir cuit es En cas de non utilisation prolong e de l appareil sortez les piles du compartiment piles Utilisez uniquement des piles identiques ou quiva lentes Remplacez toujours l ensembl...

Page 14: ...2 positionnez le n buliseur en mailles 3 bloquez le en le tournant dans le sens des aiguilles d une montre jusqu ce que les lectrodes de contact argent es pointent vers le bas Montez le r ci pient av...

Page 15: ...le couvercle 1 de l appareil 2 Remplissage du n buliseur Ouvrez le r cipient m dica ments 2 en ouvrant l trier et remplissez le r cipient m dicaments avec une solution de chlorure de sodium isoto niqu...

Page 16: ...n d viter tout risque pour votre sant Le n buliseur et les accessoires sont destin s un usage multiple Diff rentes exigences en mati re de nettoyage et de pr paration hygi nique existent pour les diff...

Page 17: ...halation sans interruption car un long intervalle entre les inhalations pourrait obstruer le n buliseur en mailles et donc affec ter l efficacit du traitement D sinfection Veuillez suivre scrupuleusem...

Page 18: ...transportez pas avec des objets humides Rangez et transportez l appareil et les accessoires l abri des rayons directs du soleil D pannage Probl mes questions Cause possible solution Le n buliseur ne...

Page 19: ...r serve de modifications techniques Les pi ces d usure sont exclues de la garantie Le num ro de s rie se trouve sur l appareil ou sur le compartiment piles Les mesures ont t effectu es avec une solut...

Page 20: ...mpl tement d charg es doivent tre mises au rebut dans des conteneurs sp ciaux ou aux points de collecte r serv s cet usage ou bien d pos es chez un revendeur d appa reils lectriques L limination des p...