Table 2
For all me equipment and me systems
Guidance and manufacturer’s declaration – electromagnetic im-
munity
Guidance and manufacturer’s declaration – electromagnetic immunity
EM34 is intended for use in the electromagnetic environment specified below. The customer or the user of EM34 should assure that it is used
in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment – guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic materi-
al, the relative humidity should be at least 30 %.
Electrical fast transient/burst
IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV input/output lines
Not applicable
Mains power quality should be that of a typical
commercial or hospital environment.
Surge IEC 61000-4-5
± 1 kV differential mode
± 2 kV common mode
Not applicable
Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short interrup-
tions and voltage variations on
power supply input lines IEC
61000-4-11
< 5 % U
T
(> 95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
< 5 % U
T
(> 95 % dip in U
T
)
for 5 sec
Not applicable
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of EM34 requires continued operation during
power mains interruptions, it is recommended
that EM34 is be powered from an uninterrupti-
ble power supply or a battery.
Power frequency (50/ 60 Hz)
magnetic field IEC 61000-4-8 3 A/m
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
NOTICE:
U
T
is the a.c. mains voltage prior to application of the test level.