Beurer GL50
21
3
Affix protective disc to needle
Place the retained protective disc flat on a hard surface. Stick the tip of the needle into the
protective disc (image 1) so the needle is covered (image 2).
Image 1
Image 2
4
Remove and dispose of lancet needle
Press the trigger again so that you can grip the shaft. Carefully remove
the lancet needle from the lancing device and dispose of the lancet in
a puncture-proof container.
Carefully dispose of all blood samples and materials that you have come
into contact with. This prevents injuries and the infection of others.
5
Replace cap
Place the cap back on the device.
5.6 Evaluating measured blood glucose values
Your blood glucose monitor can process measured values between 20 and 630 mg/dL (1.1 and
35.0 mmol/L). The “ ” warning is displayed for measured values below 20 mg/dL (1.1 mmol/L). The
“ ” warning is displayed for measured values above 630 mg/dL (35.0 mmol/L).
Warning
• If you suspect that the blood glucose results are incorrect, first repeat the test and, if applicable,
perform a functional test using control solution. Seek medical advice if dubious results persist.
• Seek medical attention immediately if your symptoms do not correspond to your measured blood
glucose values and you have followed all instructions for the Beurer GL50 blood glucose measuring
system.
• Do not ignore symptoms of too high/low blood glucose levels. Always seek medical attention!