![Berthold 55393 BA2 Operating Manual Download Page 115](http://html1.mh-extra.com/html/berthold/55393-ba2/55393-ba2_operating-manual_2739252115.webp)
LB 983 NightSHADE Operating Manual
Appendix
115
The control cable of each panel is plugged into the socket at the corresponding side of the con-
nection box (see 5.3.3).
Water cooling connects to the water supply bar located at the rear side of the dark box. The
plug at the lower side of the is a bypass plug (instead of the temperature controlled sample ta-
ble) and must be present to ensure a flow of cooling water.
External connectors for the cooling liquid are located at the rear side of the miniChiller and the
rear panel of the NightSHADE:
Tubing connector at the miniChiller and at the rear panel of the NightSHADE
When the miniChiller or any other device using liquids to cool the LED panel is connect-
ed to the NightSHADE it is not allowed to operate 100/230V devices such as
transilluminators etc. inside the dark room due to electrical safety regulations.
The preset temperature of the cooling water in the miniChiller must not be set below am-
bient temperature since this could lead to condensation on the LED panel.
The
“Artificial daylight” control must be enabled in the indiGO service backend in order to con-
trol the intensity of the LED panels.