Operation and Menu Structure
NewSonic SonoDur-R
Issue 105 06/2020
Page 35
7.8
Instrument Menu
Press the Menu button to access the instrument menu with access to all device parameters, the data
memory and the adjustment function. Some menu lists are larger as on one screen could be
displayed. The menu list can be dragged up and down to make all menu lines visible. A “slider” shows
the actual position at the right screen area (see
).
The setting options or current settings are displayed in short form below the respective menu item in
blue.
Switching to the selected submenu is invoked immediately by tapping a row. The name of the sub-
program is displayed again at the top of the status bar.
7.9
Adjustment
Here, the measuring system can be adjusted to the test material (previously frequently incorrectly
referred to as calibration) by determining an adjustment number that includes the material
properties (E-Module deviates from low-alloy steel) and is applicable for the specified test method.
Such adjustment number depends on the testing force, the probe type (hand-held measuring probe
or motor measuring probe) and on the test procedure (penetration time and measuring direction).
Saved adjustments only apply to the respective probe type.
The adjustment can be started via the sub-program "Adjust measured value" in which the
adjustment number is determined directly via an adjustment measurement directly on the test piece
or via "Adjustment number direct" and the direct input of the adjustment number without
performing an adjusting measurement, in which case all measured values will be immediately
recalculated based on the directly entered adjustment number.
7.9.1
Adjusting the measuring value
There are 3 ways to perform an adjustment:
▪
Adjustment by using the measured values of the already recorded measurement series.
▪
Recording a new measurement series.
▪
Input of an adjustment number, if such number is already known.
Figure 7.45
Figure 7.46
Figure 7.47