BV200 Plasma Viscometer Mk2/3 Customer User
Manual
Issue 9
th
Sep 2020 - Rev 1.8
Page 45 of 53
6.8. Drain Line Filter Cleaning
Remove the inline filter from the drain line between the washpot and the drain valve.
1.
Remove the 3mm (black tail) tubing and connector from the cassette cleaning syringe.
2.
Fill the syringe with 20ml of de-ionised water.
3.
Push the 4mm (green tail) tubing into the top green coloured part of the filter.
4.
Flush the entire contents of the syringe through the filter ensuring that the waste is
5.
going in to a dirty sink.
If debris remains in the filter repeat steps 3
– 5.
6.
Fluid can pushed and sucked through the blue connection on the filter, using the
7.
syringe, to remove any stubborn particles.
IMPORTANT
– Flush the cassette syringe after use to remove any contamination and
8.
debris to prevent this being injected into the cassette during cleaning.
Luer adaptor
Green coded
4 mm ø tube
Syringe
Drain
Filter
Flow
Fill the syringe
with de-ionised
water. Flush
the entire
contents of the
syringe via the
green
connector
through the
filter.