65
IRDH275_D00122_03_M_XXEN/05.2017
7. Serial interfaces
The ISOMETER®s IRDH275 and IRDH275B have differently designed serial in-
terfaces.
7.1
RS485 interface with IsoData protocol (IRDH275)
Data transmission is continuously carried out and can neither be interrupted
by the data slave station nor be influenced in any other way. This protocol
cannot be used in combination with the BMS protocol.
For data evaluation via PC or Laptop, the terminal software "IsoData" and an
interface converter of the ASCII-RS485/RS232 type is required. To obtain the
software, please contact Bender Service. Contact details are found on page 8.
Interface data:
•
RS485 interface galvanically isolated from the device electronics
•
Connection to terminal A and B
•
Maximum cable length 1200 m
•
Transmission protocol 9600 baud - 1 start bit - 1 stop bit - 8 data bit
•
After each valid measurement, the following data block is provided:
IRDH275
- RS485 and IsoData protocol
- galvanically isolated
- ASCII, unidirectional
IRDH275B
- RS485 and BMS protocol
- galvanically isolated
- ASCII, bidirectional
Summary of Contents for ISOMETER IRDH275
Page 36: ...Anschluss 36 IRDH275_D00122_03_M_XXDE 05 2017 ...
Page 40: ...Inbetriebnahme Schema dreiteilig 40 IRDH275_D00122_03_M_XXDE 05 2017 ...
Page 91: ...91 IRDH275_D00122_03_M_XXDE 05 2017 W Werkseinstellungen 73 Z Zulassungen 79 ...
Page 128: ...Connection 36 IRDH275_D00122_03_M_XXEN 05 2017 ...
Page 132: ...Commissioning flow chart threepart 40 IRDH275_D00122_03_M_XXEN 05 2017 ...
Page 144: ...Operation and setting 52 IRDH275_D00122_03_M_XXEN 05 2017 Diagram ISO SETUP ...
Page 152: ...Operation and setting 60 IRDH275_D00122_03_M_XXEN 05 2017 6 6 4 Diagram COM SETUP IRDH275B ...
Page 154: ...Operation and setting 62 IRDH275_D00122_03_M_XXEN 05 2017 6 7 2 Diagram PASSWORD ...