32
IP Panels © 2018 Bender Inc. All Rights Reserved
Installation
7.8.4
Grounding
As defined by NEC Article 517, the ground bus in the isolated power system is
the reference grounding point for the room. Refer to figure below for typical
appearance and location.
•
Connect all ground conductors from the equipment and receptacles in
the area served by the isolated power system to the panel’s ground bus.
•
If more than one isolated power system is serving the same area, bond
them together using a minimum AWG 10 insulated equipment bonding
conductor per NEC Article 517.14.
7.8.5
Accessories / Add-On Kit Wiring
Refer to the respective installation manual(s) for installing additional equip-
ment.