SECTION 6 : POST INSTALLATION  CONFIRMATION

  [ 1 ] CONFIRMATION OF POWER SUPPLY VOLTAGE

 As specified in Electrical Requirements (page 5), power supply voltage must be within the

 operable range

. (Rated 

line 

voltage ±10%)

 

Confirm the power supply voltage again before  

      

turning on 

the 

unit.

1. Open the front panel of main controller by loosening two screws on top of the controller.

2. Set the range of digital multimeter at 

3

00 VAC, connect probes of multimeter to L and N of the

3P terminal block.

Do not touch the restriction plate (refer to Fig.3-3) with the probes of multimeter during

measurement, or a short circuit occurs.

3. Confirm that the reading is 

rated line voltage ±

10%

.

NOTE  : PHOT-XIIs 

505

 x-ray can not be operated unless the power supply voltage is

 within

this range.

WARNING

[ 2 ] CONFIRMATION OF TUBE POTENTIAL COMPENSATION VALUE

 Tube potential is kept to be the constant and specified value by the feed-back control system.

   Hight voltage is converted into low voltage feed back signal by the voltage divider. The precision

   of tube potential depends on the accuracy of this voltage divider, although each divider has little

   deviation. To compensate this deviation, we prepare the compensation value for each tube head.

   Before making an exposure, check this value to be same as the value stored in the subcontroller.

1. Confirm the tube potential compensation (CP) values for 60kV and for 70kV written on the label

attached inside of the head yoke.

2. Turn on the main power switch. Keep depressing P(patient), C(cone) and kV selection switches

together until “CP.  ” is displayed in exposure time window.  This value is for 60kV and should

be the same value on the label.  If displayed value is different, press Exposure Time Adjusting

Switches (  or  ) and make the CP value to be same as the label and press patient switch to
store it

3. Press kV switch, then CP value for 70kV will be displayed.  This value should be same as the

value for 70kV written on the label.  If it is different, adjust displayed value by the Exposure

Time Adjusting Switches (  or  ) and press patient switch to store it.

[ 3 ] MA

 (

TUBE CURRENT

) ADJUSTMENT

   PHOT-XIIs 

505

 x-ray incorporates self diagnose and adjusting system to check if the tube

 current

   are within specified ranges at the beginning of exposure.

1. Keep depressing tooth selection switches T1, T4 & T5 together until “h.  ” is appeared on the

      exposure time display window

2. Wait until the display changes to be “0.

50

”.

3. Make exposure by depressing the exposure switch.

X-radiation is generated for 0.

5

 second.

4. Repeat step 2. and 3. until “Fin” is displayed.  This self diagnose and adjustment is automatically

done for 3mA

 and 

6mA.

WARNING

24

Summary of Contents for PHOT-X IIs 505

Page 1: ...nstallation assembly calibration and certification procedures for BELMONT PHOT X IIs 505 dental x ray The instructions contained in this book should be thoroughly read and understood by dealer service...

Page 2: ......

Page 3: ...IDENTIFICATION AND OPERATION 1 MAJOR COMPONENTS AND CONTROL DENTIFICATION 17 2 FUNCTION OF CONTROLS 18 3 OPERATING PROCEDURES 21 4 ESTIMATED AIR KERMA 21 5 OPTIONAL HAND EXPOSURE SWITCH 21 6 ERROR CO...

Page 4: ...utput 60 kV 70 kV 3 mA 6 mA 3 mA 6 mA a Distal end of regular cone 4 6 9 1 5 9 11 8 mGy sec 40 b Distal end of long cone 2 0 4 1 2 6 5 2 mGy sec 40 Data obtained by direct measurement in the useful be...

Page 5: ...Maximum Reach 185 1 15 PHOT X IIs Sub Controller 34 mm Front to Back Box Thickness P 150 178 211 7 7 23 100 6 1 130 1 16 0 1 29 4 85 2 31 50 973 1023 1243 Short Cone Long Cone Cabinet Frame for X ray...

Page 6: ...both B Duty cycle A cool down interval of 30 seconds or more must be allowed between each 1 second exposure a 15 second cool down must be allowed between each 0 5 second exposure This will avoid the a...

Page 7: ...n device Use a flexible cable approved by CEE 13 52 or 53 consists of 0 75 mm2 or 1 mm2 conductors Diameter of the sheath of cable should be 6 7 5 mm diameter Line voltage regulation should be within...

Page 8: ...wing radiation requirements concerning the operator s positioning must be considered The operater must 1 have full view of the patient 2 have means for audio and visual communications with the patient...

Page 9: ...ENTS 2 UNPACKING Unpack the entire contents of the shipping carton Included within the shipping carton are Hardware Quantity Head with Regular Cone 1 Main Controller 1 Screw for chassis M4 x 10 mm 4 S...

Page 10: ...ring matches to the access hole of wall plate template 2 Mark the hole locations for the arm mounting bracket lag screws 3 Use a 3 16 5mm dia drill to make a pilot hole approximately 50mm deep for eac...

Page 11: ...nting bracket Fig 3 5 Insert the cable of horizontal arm into the hole and mount to the arm mounting bracket as shown in Fig 3 5 2 Insert two retaining bolts into the upper threaded holes of the arm m...

Page 12: ...de toward the wall to close Fig 3 9A 2 Install two M3 x 8mm screws into the top of the cover and confirm that the cover is securely attached Fig 3 9B M3x8 Screw Front Cover Front Cover Front Cover Cha...

Page 13: ...3A Note Two ring terminals should not protrude from a chassis plate Fig 3 13A 5 Secure the wires from the balance arm to the bottom cover with the nylon cable clamp to prevent damage from twisting Fig...

Page 14: ...n the band 1 Remove the arm collar screw M4 x 8mm from the arm collar Slide the arm collar upward and temporarily hold it in position with adhesive tape Fig 3 15 2 Open the yoke inside cover of x ray...

Page 15: ...ith the nylon cable clamp Fig 3 18 7 Reattach the yoke inside cover with the screw M4 x 15mm Before closing the cover note the CP values on the CP values label inside of the yoke Fig 3 16 8 Remove arm...

Page 16: ...ler Fig 3 21 Miswiring causes permanent damage to both timer PC board and power PC board 5 If wire length is too long push it back into the access hole of the wall This will prevent mechanical damage...

Page 17: ...ith the tapping screw 3 x 8mm Fig 3 25 Fig 3 23 Hand Exposure Switch Kit Hand Exposure Switch Hook Tapping Screw 3 x 8 Fig 3 24 Connecting Hand Exposure Switch Timer PC Board Front Panel CN3 CN2 CN3 H...

Page 18: ...light tendencies to drift can be corrected by tightening brake screw in horizontal arm Do not tighten beyond what is required to prevent drift 2 BALANCE ARM TENSION ADJUSTMENT 1 Place the balance arm...

Page 19: ...tion Switch 2 Ready Light 11 Film Speed Selection Switch 3 ExposureTimeAdjusting Switch Down 12 Digital Imaging Switch 4 ExposureTimeAdjusting Switch Up 13 kV Selection Switch 5 Tooth Selection Switch...

Page 20: ...GHSUHVVHG PRUH WKDQ VHF XQLW JRHV LQWR RFN 0RGH Q ORFN PRGH WKH RQO IXQFWLRQDO VZLWFK LV WKH SRZHU VZLWFK 7R H LW IURP WKH ORFN PRGH GHSUHVV WKH 7 VZLWFh PRUH WKDQ VHF DJDLQ RQH 7 SH 6HOHFWLRQ 6ZLWFK...

Page 21: ...Q WKH SRVXUH 7LPH LVSOD LQGRZ LWK WKH IDFWRU VSHHG VHWWLQJ G 0 WKH H SRVXUH WLPH EHFRPHV KDOI RI 10 VHWWLQJ As the sensitivity is different according to each manufacturer of digital imaging sensors th...

Page 22: ...13 0 16 0 18 0 22 0 32 0 22 0 25 0 32 0 36 0 50 0 28 0 32 0 40 0 45 0 63 3 0 18 0 22 0 28 0 32 0 45 0 32 0 36 0 45 0 50 0 71 0 40 0 45 0 56 0 63 0 90 6 0 09 0 11 0 13 0 16 0 22 0 16 0 18 0 22 0 25 0...

Page 23: ...ures when the unit is not in XVH 127 I WKH XQLW OHIW RYHU PLQ ZLWKRXW EHLQJ RSHUDWHG DQG WKH 0DLQ 3RZHU 6ZLWFK LV NHSW RQ ILJXUH UXQV WKURXJK WKH SRVXUH 7LPH LVSOD LQGRZ 7KLV GRHV QRW PHDQ WKDW PDOIXQ...

Page 24: ...re than 4 5 mA at 3mA setting or more than 9 mA at 6 mA setting Setting for pre heating time is out of range Tube current is detected during pre heating period Tube current is detected when main power...

Page 25: ...QLQJ ZLWKLQ WKH PDQXIDFWXUH V VSHFLILFDWLRQV DQG UHPDLQV LQ FRPSOLDQFH ZLWK WKH 6WDQGDUG W LV UHVSRQVLELOLW RI WKH RZQHU RI WKH XQLW WR VHH WKDW WKHVH PDLQWHQDQFH FKHFNV DUH correctly performed Any pa...

Page 26: ...re making an exposure check this value to be same as the value stored in the subcontroller 1 Confirm the tube potential compensation CP values for 60kV and for 70kV written on the label attached insid...

Page 27: ...g light is located on the front panel of the sub controller 1 Make an exposure and confirm that the warning light illuminates during the exposure 6 CONFIRMATION OF LINE VOLTAGE REGULATION 1 Make sure...

Page 28: ...ase film speed number until desired number for a setting is displayed 3 To change the b setting from the factory default F 05 push F switch momentarily until the b light illuminates By depressing or s...

Page 29: ...conds Release the switches if the ready light starts to flash 2 bu 2 will be displayed in exposure time display window 3 By depressing either or switch display changes to bu 0 4 Press P switch Patient...

Page 30: ...GE BLUE BROWN BLACK GRAY WHITE ORANGE PURPLE YELLOW BLUE BROWN GRAY RED WHITE BLUE GRAY RED WHITE YELLOW BROWN BLACK BROWN GREEN YELLOW RED BLUE YELLOW BLACK 3 1 2 SWITCH EXP 1 2 3 4 TIMER PC BOARD RE...

Page 31: ...racket to the wall plate are loose find the reason why bolts became loose take counter measure that prevent bolts become loose 6 Dosimetry Save the image that was taken under appropriate conditions as...

Page 32: ...ats Mats Mats Mats APPENDIX 3 PARTS IDENTIFICATION 1 CONTROLLER ASSEMBLY C T5 T2 T3 T4 kV 60 mA INC CUS PM CUS PM MLR OCC MLR BW 31 33 34 39 32 44 27 35 29 30 28 43 Mats 36 11 11 40 41 45 46 1 2 38 3...

Page 33: ...r PC Board 240V Timer PC Board 110V Timer PC Board 100V Timer PC Board 120V Timer PC Board 220V Timer PC Board 230V 31 ECQR68B0 Hook for Hand Exposure Switch Option 32 Hand Exposure Switch Hand Exposu...

Page 34: ...e Inside Cover Screw M4 x 15 1 5 Head Housing Cover Set 1 set 6 Yoke Side Cap 1 7 Yoke Side Cap Screw 3 x 8 1 8 Lock Ring 1 9 Regular Cone 1 10 Long Cone Option 1 11 Rectangular Collimator Option 1 12...

Page 35: ...NOTE...

Page 36: ...Book No 1A0GDSG0 Printed in Japan 2017 04 2 1 1 Higashishinsaibashi Chuo ku Osaka 542 0083 Japan TEL 81 6 6213 5945 TELEFAX 81 6 6212 3680 TAKARA BELMONT CORPORATION NOTE R...

Reviews: