-13-
4. Insert the shaft of the balance arm into the
head yoke, and while holding the head in
position, insert the head key securely into the
retaining groove. (
Fig.3-17
)
5. Remove adhesive tape and slide the arm collar
downward. Fix it in place with the arm collar
screw.
Remove the UP-mark from the arm collar
.
(
Fig.3-17
)
6. Loosen the (M5 x 10mm) screw and remove
the nylon cable clamp from the yoke housing.
Place cable clamp on the balance arm cable.
Connect the 10P connectors, and then attach
the balance arm cable to the yoke housing
with the nylon cable clamp. (
Fig.3-18
)
7.
Reattach the yoke inside cover with the screw
(M
4
x
15
mm). Before closing the cover
,
note
the CP values on the CP values label
inside
of the yoke. (
Fig.3-16
)
8
. Remove arm holding band.
Fig.3-17
X-ray Head Installation
CAUTION !
DO NOT RELEASE
THIS BAND UNTIL
X-RAY HEAD IS
INSTALLED
CAUTION !
DO NOT RELEASE
THIS BAND UNTIL
X-RAY HEAD IS
INSTALLED
Arm Holding
Band
Caution Tag
Arm Collar
Screw (M4 x 8)
Arm Collar
UP-Mark
Head Key
Stopper Ring
UP
Yoke
Head
Fig.3-18
Connection 10P Connector in Yoke
3. Making sure the s
t
opper ring is placed on the
yoke, insert the wiring from the balance arm
assembly through the head shaft and into the
yoke. (
Fig.3-17
)
Nylon
Clamp
(M5 x10) Screw
10P
Connector
Yoke
Housing