Operation
64
METPOINT
®
OCV
Example setting 2
If the residual oil content values indicated in the METPOINT OCV
®
MEASUREMENT mode lie in the range between
≥
0.1 mg/m³ and
≤
1.0 mg/m³, please enter "
0.01"
into the blue field and confirm with OK.
Scaling
0.1 mA corresponds to
0.01
mg/m³
Scaling initial value
4.0 mA corresponds to
0.00
mg/m³
Scaling final value (Full-Scale)
20 mA corresponds to
1.60
mg/m³
Example setting 3
If the residual oil content values indicated in the METPOINT OCV
®
MEASUREMENT mode lie in the range between
≥
1.0 mg/m³ and
≤
5.0 mg/m³, please enter "
0.1"
into the blue field and confirm with OK.
Scaling
0.1 mA corresponds to
0.1
mg/m³
Scaling initial value
4.0 mA corresponds to
0.0
mg/m³
Scaling final value (Full-Scale)
20 mA corresponds to
16
mg/m³
For all other measured-value ranges, the 4-20 mA signal can be scaled using the same steps.