6.  MAINTENANCE

6.1 Fault finding during commissioning

If a BA644DF-F fails to function during
commissioning the following procedure should be
followed:

Symptom

Cause

Check:

No Display

Instrument not

correctly

connected or

powered.

9 to 32V between

terminals 1 & 2.

When configured

as a listener

displays

‘no ConF’

No inputs are

enabled.

Configuration of

each input.

When configured

as a listener

displays

‘no dAtA’

No data being

received with the

current

configuration.

Addresses and

Index are correct.

That data has

DS-33 format.

Display shows

‘9.9.9.9.9’ with all

decimal points

flashing; all

bargraph

segments

activated and

bargraph scale

flashing.

Value

over-range

Configuration.

and

decimal point

position.

Display shows

‘-9.9.9.9.9’ with

all decimal points

flashing; no

bargraph

segments

activated and

bargraph scale

flashing.

Value

under-range

Configuration

And

Decimal point

position.

Display

alternates

between value

and the word

‘bAd’.  Bargraph

flashes.

Status of fieldbus

variable has a

quality of ‘BAD’

or a fault state is

active.

Fieldbus

configuration.

Bargraph scale

flashes.

Process variable

is outside the

limits defined for

the bargraph.

Bargraph

configuration.

All display

segments
activated.

Display is

initialising.

This is normal

operation, after a

few seconds the

model number

followed by the

firmware version
will be displayed

prior to entering

the display

mode.

6.2  Fault finding after commissioning

ENSURE PLANT SAFETY BEFORE

STARTING MAINTENANCE

If a BA644DF-F fails after it has been functioning
correctly, the table shown in section 6.1 may help
to identify the cause of the failure.

If this procedure does not reveal the cause of the
fault, it is recommended that the instrument is
replaced.

6.3  Servicing

We recommend that faulty BA644DF-F
FOUNDATION™ fieldbus indicators are returned
to BEKA associates or to our local agent for repair.

6.4 Routine maintenance

The mechanical and electrical condition of the
instrument should be regularly checked.  Initially
annual inspections are recommended, but the
inspection frequency should be adjusted to suit the
environmental conditions.

6.5  Guarantee

Instruments which fail within the guarantee period
should be returned to BEKA associates or our local
agent.  It is helpful if a brief description of the fault
symptoms is provided.

6.6  Customer comments

BEKA associates is always pleased to receive
comments from customers about our products and
services.  All communications are acknowledged
and whenever possible, suggestions are
implemented.

Summary of Contents for FOUNDATION BA644DF-F

Page 1: ...Issue 4 4th March 2014 BA644DF F FOUNDATION fieldbus Field mounting 8 variable Fieldbus Indicator Issue 4...

Page 2: ...display parameters 5 3 6 Enable selected input 5 3 7 Enter Device Address 5 3 8 Enter Variable Identity 5 3 9 Decimal point position 5 3 10 Configuration of bargraph 5 3 11 Scale numeric display 5 3...

Page 3: ...protocol is shown on a label inside the terminal cover The F order code suffix also indicates the protocol but is not shown on the instrument identification label There is an alternative version of t...

Page 4: ...arth terminal should be connected to the potential equalising network The BA644DF F enclosure is supplied with a bonding plate to ensure electrical continuity between the three conduit cable entries 4...

Page 5: ...5 Fig 3 Dimensions and terminal connections...

Page 6: ...nditioned as a fieldbus listener see section 5 3 13 to change it to a fieldbus node 5 2 1 Reset the indicator to node factory defaults rESEt The BA644DF F may be reset to the factory defaults that are...

Page 7: ...the P button causes the BA644DF F indicator to search the fieldbus segment to which it is connected for all publishers Fig 5 shows how the indicator displays the hexadecimal Device Address and Variabl...

Page 8: ...vice Address and Variable Identity Parameters can not be changed from within this function See section 5 3 4 Edit Enter addresses and display parameters Contains six sub functions that enable known fi...

Page 9: ...9...

Page 10: ...isplayed pressing E will allocate the selected process variable to the selected input and return the display to the Device Address and Variable Identity from which another process variable may be sele...

Page 11: ...function pressing P again will reveal the existing bargraph justification The or button will scroll through the four options when the required justification is displayed pressing E will enter the sele...

Page 12: ...ration menu using the or button To enter the function press P which will show that the instrument is configured as a fieldbus listener to change to a fieldbus node press the or button which will toggl...

Page 13: ...nfiguration All display segments activated Display is initialising This is normal operation after a few seconds the model number followed by the firmware version will be displayed prior to entering th...

Page 14: ...After adding the required legends or fitting a new pre printed self adhesive escutcheon ensure that the gasket is correctly positioned before reassembling the instrument enclosure 7 2 Tag plate The BA...

Reviews: