background image

15

stored in memory, and will then return to the
'SPAn' prompt.  Finally press E again to
return to the main menu.

This function may also be used when
calibrating a complete loop including the flow
meter.

6.8 Calibration of the rate display using
      internal references:   SEt

This function enables the zero and span of
the rate display to be adjusted without the
need for an accurate external current
source, or for disconnection from the
4/20mA loop.  The accuracy of this method
depends upon the accuracy of the internal
references which should be regularly
calibrated as described in section 6.9.2

To calibrate the rate display select 'SEt' from
the main menu and press P.  The instrument
will display 'ZErO', pressing P again will
reveal the current display at 4mA.  For flow
applications this will usually be zero.  Each
digit of the rate display can be changed by
pressing the Up or Down buttons.  When the
first digit is correct pressing P will transfer
control to the next digit.  When the least
significant digit has been adjusted, press E
to return to the 'ZErO' prompt which
completes the adjustment.

To adjust the rate display at 20mA, press the
Up button which will cause the instrument to
display 'SPAn'.  Pressing again will reveal
the existing full scale rate display.   Each
digit can be changed by pressing the Up or
Down buttons. When the first digit is correct,
pressing  P will transfer control to the next
digit. The most significant digit can only be
adjusted by rolling-over the previous digit.
When all the digits have been adjusted
press  E to return to the 'SPAn' prompt,
followed by E to return to the main menu.

6.9  Conditioning sub-menu:   Cond

This sub-menu allows the mains (line)
frequency at which the instrument has
maximum ac rejection to be selected and the
two internal references to be calibrated. To
minimise the possibility of accidental
adjustment these functions are contained in
a protected sub-menu.  To gain access to
this sub-menu select 'Cond' from the main
menu and press P which will cause the
instrument to display '0000'.   Using the Up,
and Down buttons and the P button to move
to the next digit, enter the password 'SurE'.
'S' is entered as '5'.  Pressing E will then
give access to the sub-menu.

6.9.1  AC rejection:   FrE

WARNING

If the mains filter frequency is
changed, the rate display and the
internal references (if used) must be
recalibrated.

To provide maximum low frequency rejection
the internal digital filter may be set to
operate at 50 or 60Hz to correspond with the
local mains (line) frequency.   To change the
frequency select 'FrE' from the 'Cond' sub-
menu and press P which will reveal the
current setting.  The setting can be changed
by pressing the Up or Down buttons followed
by the E button to return to the sub-menu.

6.9.2  Calibration of internal references:
          rEF

The BA358C contains two references
representing 4 and 20mA.  These are used
in the 'SEt' function that enables the rate
display to be calibrated without an external
current calibrator.  They are also used in the
'C--P' function when the P push-button is
programmed to display the input current in
the operating mode.   If either of these two
functions are used, the internal references
should be routinely calibrated against an
external 4/20mA calibrator connected to the
input terminals of the instrument.   See
section 8.4

To calibrate the internal references select
'rEF' from the sub-menu and press P  which
will result in a '.004A' prompt being
displayed. Adjust the external current
calibrator to 4.000mA  and again press P.
The instrument will display 'Ent' while the
4mA reference is being updated and will
then return to the '.004A' prompt.

To re-calibrate the 20mA internal reference,
press the Up button which will cause the
instrument to display '.020A'. Adjust the
external current calibrator to 20.000mA  and
again press P.  The BA385C will display
'Ent' while the 20mA reference is being
updated and will then return to the '.020A'
prompt.  This completes calibration of the
internal references, two operations of the E
button will return the instrument to the 'Cond'
prompt in the main menu.

6.10  Resetting grand total:   CLr. Gtot

The grand total is a separate sixteen digit
counter which duplicates the total display,
but is not zeroed when the total display is
reset to zero.  The grand total may be

Summary of Contents for BA358C

Page 1: ...October 2010 BA358C Intrinsically safe 4 20mA loop powered panel mounting rate totaliser Issue 4 Certification information label Rate display Annunciators for optional alarms Rotating flow indicator T...

Page 2: ...1 Timebase 6 12 Function of the P push button 6 13 Total scale factor 6 14 Percentage clip off 6 15 Local reset of total display 6 16 Security code 6 17 Over and under range 7 Calibration Examples 7 1...

Page 3: ...sed diode D1 The voltage developed across D1 which is relatively constant is multiplied by a switch mode power supply and used to power the instrument The voltage developed across R1 which is proporti...

Page 4: ...eds 4mA 3 INTRINSIC SAFETY CERTIFICATION 3 1 ATEX certificate The BA358C has been issued with an EC Type Examination Certificate by Notified Body Intertek Testing Services ITS formerly ERA Technology...

Page 5: ...Zones gas groups and T rating The BA358C has been certified to show compliance with the requirements for Group II Category 1G equipment EEx ia IIC T5 When connected to a suitable system the instrumen...

Page 6: ...dication of rate and total flow within a hazardous area The type of interface is not critical either a Zener barrier or a galvanic isolator may be used providing that Uo Io and Po are not exceeded If...

Page 7: ...esetting from the safe area via a single channel diode return Zener barrier A certified intrinsically safe relay or certified opto isolator may also be used Fig 3B shows how the total display may be r...

Page 8: ...lective displays there is little degradation of contrast when viewed from above the centre line but slight degradation may be noticeable when the instrument is fitted with a backlight Fig 4 BA358C dim...

Page 9: ...in the rate display is defined by the d p function To simplify calibration the BA358C uses floating point arithmetic moving the rate display decimal point position will therefore affect totalisation T...

Page 10: ...of 0 00 with 4mA input and 100 00 with 20mA input The total scale factor will be set to 1 0000 with a timebase of flow per second The instrument can easily be recalibrated on site 6 3 Summary of progr...

Page 11: ...input current in milliamps or the input as a percentage when the P push button is operated in the display mode Select PC for of span and 4 20 to display current See section 6 12 Display Description o...

Page 12: ...cimal point clip off to adjust value Press or between 0 and 10 or and pressing Define Security Enter by Code to move to next digit Local reset or Fig 7B Programme Structure CLr Gtot E E E E E E E E P...

Page 13: ...or and Similarly set the required full scale display accurate 20mA input current scale display using an Similarly set the required full to move to next digit pressing or and set required zero display...

Page 14: ...s Operating the P push button will toggle control between the two displays When both decimal points have been correctly positioned press E to return to the main menu If the application requires flow r...

Page 15: ...ay 0000 Using the Up and Down buttons and the P button to move to the next digit enter the password SurE S is entered as 5 Pressing E will then give access to the sub menu 6 9 1 AC rejection FrE WARNI...

Page 16: ...our Tb 3600 3600 Select the required divisor and press E to return to the main menu 6 12 Function of the P push button C P This parameter defines the function of the P push button when the BA358C is i...

Page 17: ...lect CLIP OFF from the main menu and press P to reveal the current setting This clip off threshold is shown as a percentage of total flow and may be adjusted in one percent increments using the Up or...

Page 18: ...ction 7 1 describes the preferred method of calibration using an external traceable 4 20mA calibrator Section 7 2 contains an example of calibration using the BA358C internal 4 and 20mA references whi...

Page 19: ...git of the total display flashing Press the Up or Down push button until the second least significant digit and following decimal point of the total display flash This gives the required total display...

Page 20: ...to the main menu by pressing E Step 12 Reset the grand total to zero To reset the grand total to zero select CLr Gtot from the main menu and press P which will cause the instrument to display Clr no T...

Page 21: ...multaneously pressing P and E Assuming a security code has not already been entered the instrument will respond by displaying root which is the first function in the main menu Steps Perform as describ...

Page 22: ...terminals 1 3 No power supply Confirm that a 4 20mA current is flowing in the loop Unstable rate display 4 20mA current is noisy or the flow signal is unstable Eliminate source of electrical noise Red...

Page 23: ...ons available for a rate alarm and shows which are fail safe i e output is in the alarm condition open when the 4 20mA input current is zero Similarly when programmed as a total alarm the output shoul...

Page 24: ...nt capacitance and inductance of each BA358C alarm output is Ci 40nF Li 20 H The maximum permissible cable parameters for the alarm circuit will be the figures specified for the barrier or isolator le...

Page 25: ...veal the current setting The function can be changed by pressing the Up or Down button followed by the E button to return to the alarm menu 9 3 5 Type of alarm tYPE This function conditions the alarm...

Page 26: ...Down button followed by the E button to return to the alarm menu 9 3 9 Hysteresis HStr Hysteresis is only available on rate alarms During programming hysteresis is shown in the units of rate the BA358...

Page 27: ...This direct access menu is enabled and a separate security code entered from the AcSP function in the main programme menu To change the menu parameters select AcSP from the programme menu and press P...

Page 28: ...is segregated from the measuring circuit and has been certified as a separate intrinsically safe circuit but it does not comply with the requirements for simple apparatus Systems certificates for use...

Page 29: ...ere the instrument is likely to be damaged by impact an additional front cover is available The BA398 is a robust transparent cover which is secured to the panel by six screws and sealed with a neopre...

Page 30: ...missioning 8 1 after commissioning 8 2 Flow indicator 2 2 Flow transmitter loops 4 1 Galvanic isolators 4 0 Gas groups 3 4 Grand total 2 1 clear 6 3 6 10 Guarantee 8 5 HI LO 9 3 3 9 3 7 HStr 9 3 3 9 3...

Reviews: