Appendix
Operating manual AVERO Phoenix
V2.1 E
Page 30 of 35
Guidance and manufacturer’s declaration
- electromagnetic immunity (Table 202)
The AVERO Phoenix been designed for use in the hereafter listed ELECTROMAGNETIC
ENVIRONMENTS. The customer or the user of the AVERO Phoenix must ensure that the appliance is used
in such environment.
Immunity
testing
IEC 60601- Test level
Compliance
level
Electromagnetic
environment guidance
Discharging of static
electricity (ESD) to
IEC 61000-4-2
± 6kV contact discharge
± 8kV air discharge
± 6kV contact discharge
± 8kV air discharge
The floor must be in
wood, concrete or
ceramic tiles. In case of
floors in synthetic
material, the relevant air
humidity must be at
least 30%.
Rapid transient
interference
pulses/burst
IEC 61000-4-4
± 2 kV for
power supply cables
± 1 kV for
input/
output cables
± 2 kV for
power supply cables
not applicable to input/
output cables
The quality of the
supply voltage should
match that of a typical
business or hospital
environment.
Overvoltage
IEC 61000-4-5
±/1 kV cable against
cable
±/2 kV cable against
ground connection
±/1 kV cable against
cable
±/2 kV cable against
ground connection
The quality of the
supply voltage should
match that of a typical
business or hospital
environment.
Voltage drops,
short interruptions
and
voltage fluctuations
in the power supply
input cables
IEC 61000-4-11
<5 % U
T
(>95 % drop of U
T
)
for 0.5 period
<40 % U
T
(>60 % drop of U
T
)
for 5 periods
<70 % U
T
(>30 % drop of U
T
)
for 25 periods
<5 % U
T
(>95 % drop of U
T
)
for 5 s
<5 % U
T
(>95 % drop of U
T
)
for 0.5 period
<40 % U
T
(>60 % drop of U
T
) for
5 periods
<70 % U
T
(>30 % drop of U
T
) for
25 periods
<5 % U
T
(>95 % drop of U
T
)
for 5s
The quality of the
supply voltage should
match that of a typical
business or hospital
environment.
Current frequency
(50/60 Hz)
Magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields should
be at levels
characteristic of a typical
location in a typical
business or hospital
environment.
CAUTION
U
T
is the mains AC voltage before the application of the test level.