13.7 Electromagnetic Compatibility 

Electrical medical equipment is subject to special precautionary measures with regard 
to EMC and must be installed and operated in accordance with the EMC instructions 
included in the accompanying documents. 
For the devices and systems from BEKA Hospitec GmbH, no special measures must 
be observed. 
Portable and mobile HF-communications equipment can interfere with electrical 
medical equipment. 

Guidance and manufacturer’s declaration 

- electromagnetic immunity (Table 201)

 

The AVERO Phoenix been designed for use in the hereafter listed ELECTROMAGNETIC 
ENVIRONMENTS. The customer or the user of the AVERO Phoenix must ensure that the 
appliance is used in such environment. 

Emission measurements 

Compliance 

Electromagnetic environment - 
guidelines 

High-frequency (HF) 
emissions to 
CISPR 11 

Group 1 

The AVERO Phoenix uses HF radiation 
exclusively for internal functions. 
Therefore, the HF radiation of the 
device is very low and any interference 
with adjacent electrical equipment is 
unlikely. 

High-frequency (HF) 
emissions to 
CISPR 11 

Class B 

AVERO Phoenix is intended for use in 
any type of facility including living 
quarters and those that are directly 
connected to a public mains network 
that supplies residential buildings and 
buildings used for domestic purposes. 
 

Harmonics to  
IEC 61000-3-2 

Class A 

Voltage fluctuations/ 
flicker to  
IEC 61000-3-3 

Compliant 

 

 

 

Summary of Contents for AVERO Phoenix

Page 1: ...AVERO Phoenix Lifting bath tub Version 2 1 English Subject to technical modifications 2020 11 17 Operating manual Original operating manual...

Page 2: ...List of used safety instructions 8 4 5 Warnings 9 5 Transport 10 5 1 Unpacking the product 10 5 1 1 Removing the cardboard 10 5 1 2 Loosening the product from the pallet 10 6 Installation 11 6 1 Elec...

Page 3: ...sposal 20 11 1 Disposal of the packaging material 20 11 2 Disposal of the product 20 12 Troubleshooting After sales service 21 12 1 Troubleshooting help 21 13 Appendix 23 13 1 Technical data 23 13 2 D...

Page 4: ...ured and tested according to stringent quality criteria 1 2 Manufacturer s address BEKA Hospitec GmbH Am R benmorgen 3 35582 Wetzlar Phone 49 0 641 9 22 22 0 Fax 49 0 641 9 22 22 20 info beka hospitec...

Page 5: ...ll as the national regulations laws and prescriptions related to work safety and accident prevention The manufacturer of the product is only responsible for the safety and the reliability of the produ...

Page 6: ...bly of individual components of the product Please read the operating manual and the safety instructions before starting to use the product Keep the operating manual near the device for future referen...

Page 7: ...iptions 0100 and 0100 710 However have a specialist company check the compliance of your electrical installation with the applicable prescriptions prior to operating and using the product This require...

Page 8: ...B to DIN EN 60601 1 Special waste no household waste The device and the packaging materials never must be disposed of in the domestic waste stream CE label in accordance with the EC Directive on Medic...

Page 9: ...vice in the immediate vicinity of the device is strictly forbidden Mobile telephony could also cause interference Route the mains and the connecting cable so that they cannot be damaged Damaged mains...

Page 10: ...ter in the cardboard 5 1 1 Removing the cardboard Proceed in the following way to remove the cardboard Cut the strap with the cutter knife Remove the strap Lift the cardboard up to remove it and put i...

Page 11: ...of the bath tub must be executed in accordance with the technical drawings issued by BEKA Hospitec either the guidelines for the set up and layout of wet areas and the current VDE standards All suppl...

Page 12: ...accordance with the local water supply prior to the first use Please consult the attached user manual to this effect The settings must be documented NOTE In order to ensure a safe functioning and oper...

Page 13: ...Sound system optional 7 Eccentric plug and tub inlet 3 Hand shower 8 Thermostatic mixer with scalding protection 4 Tub handle 9 Aroma system optional 5 Head neck support optional N Description N Descr...

Page 14: ...height of the bath tub can be individually adjusted If you press the UP button the bath travels up If you press the DOWN button the bath travels down 7 1 2 The hand shower The hand shower is switched...

Page 15: ...y Once the second filling height is reached the bath tub can no longer be automatically filled In this case you can add additional water to the tub by manually pressing button 9 Note Supervision of th...

Page 16: ...the DIS button to switch the disinfection on If the disinfection LED is permanently yellow the disinfection is activated Press again on the DIS button to switch off the disinfection system The disinfe...

Page 17: ...bath tub Wheelchair users can be easily lifted into the AVERO Phoenix lifting bath tub by means of mobile lift systems ceiling lifts etc The large opening underneath the bath tub provides for direct a...

Page 18: ...lvents must be used to clean the keypad Please remember that all warranty claims regarding surface damage will not be accepted if aggressive cleaning agents are anyhow used 9 2 Disinfecting the liftin...

Page 19: ...loyer s liability insurance association on stationary equipment which is used in special locations or installations an annual check to the DGUV German Statutory Accident Insurance Association Prescrip...

Page 20: ...nents must be recycled in accordance with the WEEE 11 2 Disposal of the product The expected service life of the lifting bath tub is approx 8 years At the end of the product s lifetime contact your BE...

Page 21: ...o not lit Check if the fuse and the fault current circuit breaker are activated Mains voltage fails a Check fuse b Check the cables for damages No water a Check the shut off valves of the cold and hot...

Page 22: ...pa system fails during operation a The air spa system has been used for over 15 minutes Let the air spa system cool down b Overheating due to too high an ambient temperature Let the air spa system coo...

Page 23: ...kg Fill volume 190 without patient approx 220 l Fill volume 170 without patient approx 190 l Patient s weight max 220 kg SWL safe workload 300 kg patient and water Applied part Type B Protection class...

Page 24: ...Appendix Operating manual AVERO Phoenix V2 1 E Page 24 of 35 13 2 Dimensions of the AVERO Phoenix 190...

Page 25: ...Appendix Operating manual AVERO Phoenix V2 1 E Page 25 of 35 13 3 Dimensions of the AVERO Phoenix 170...

Page 26: ...e Installed connection Sanitary Electric Warm water connection set female threading tile even Assemble hexagon nipple elbow 90 ball valve strainer and rosette see details illustration Pos 1 und 2 Cold...

Page 27: ...Appendix Operating manual AVERO Phoenix V2 1 E Page 27 of 35 13 5 Declaration of Conformity...

Page 28: ...150 Aroma system 910150587 Emergency function at power failure K9103300 Nina hand shower K9100363 Shower hose L 2 3m K9100951 Disinfectant shower yellow K9100351 Shower hose for disinfection yellow L...

Page 29: ...r listed ELECTROMAGNETIC ENVIRONMENTS The customer or the user of the AVERO Phoenix must ensure that the appliance is used in such environment Emission measurements Compliance Electromagnetic environm...

Page 30: ...quality of the supply voltage should match that of a typical business or hospital environment Overvoltage IEC 61000 4 5 1 kV cable against cable 2 kV cable against ground connection 1 kV cable against...

Page 31: ...h of fixed HF transmitters as determined by an electromagnetic field survey a should be less than the COMPLIANCE LEVEL in each frequency range b In the vicinity of equipment marked with the following...

Page 32: ...distance depending on the transmitting frequency in m 150 kHz to 800 MHz d 0 35 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 1 1 10 100 0 04 0 11 0 35 1 1 3 5 0 12 0 38 1 2 3 8 12 0...

Page 33: ...morgen 3 35582 Wetzlar Serial number _____________________________________________ Date of purchase _____________________________________________ Site _____________________________________________ Che...

Page 34: ...Appendix Operating manual AVERO Phoenix V2 1 E Page 34 of 35 Periodic inspection repair DGUV 3 safety check etc Type of inspection Date Result Measure Signature...

Page 35: ...BEKA Hospitec GmbH Am R benmorgen 3 35582 Wetzlar Germany Phone Fax 49 0 6 41 9 22 20 0...

Reviews: