37
If the final displayed value is less than 45ppm or higher than 55ppm, stop the test and perform a
calibration, following the instructions below.
ToxCOdata™ software – connecting to the PC
Place one end of the connection lead into the USB
socket on the top of the ToxCO®.
Connect the other end to the USB port on the PC.
Before starting the software, ensure that the ToxCO® is connected to the PC and switched on.
Double click the ToxCOdata™ icon on the PC to start the program. Refer to the supplied
documentation for how to operate ToxCOdata™.