26
Dispose of according to WEEE
Serial number
Manufacture by and date
Bedfont® logo
Environment
The ToxCO® complies with the directive EN60601-1-2:2015 4th edition electromagnetic
compatibility.
Electromagnetic immunity:
The ToxCO® complies with the directive EN60601-1-2 electromagnetic compatibility but can be
affected by cellular phones and by electromagnetic interference exceeding the levels specified in
EN55011:2007 Class B.
Guidance and manufacturer’s declaration: Electromagnetic immunity
The ToxCO® is intended for use in the electromagnetic environment specified below. The customer or the
user of the
ToxCO® should ensure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±8kV contact
±15kV air
±8kV contact
±15kV air
Floor should be
wood, concrete or
ceramic floor tile. If
floors are covered
with synthetic
material the relative
humidity
should be
at least 30%.
Electrical fast
transient/burst
IEC 61000-4-5
–
–
–
Surge
IEC61000-4-5
–
–
–
Voltage dips,
short
interruptions
and voltage
variations on
power supply.
IEC 61000-4-11
–
–
–
Power
frequency
(50/60Hz)
Magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location
environment.