![Beckman Coulter TQ-Prep Reference Download Page 68](http://html.mh-extra.com/html/beckman-coulter/tq-prep/tq-prep_reference_2719765068.webp)
PN 4237395AA
B-13
OPERATOR CERTIFICATION PROCEDURE
TQ-PREP WORKSTATION PRECISION PROCEDURE
B
10
When processing is done, remove the
carousel and assay the samples for
lymphocytes on a flow cytometer per
the instrument’s manual.
11
Acceptable results:
a.
The histograms must not show
unlysed erythrocytes.
b.
All 32 tubes must recover the
same percent lymphocytes
±2% of the mean using the
same analysis region.
Summary of Contents for TQ-Prep
Page 4: ...PN 4237395AA iv REVISION STATUS ...
Page 8: ...PN 4237395AA viii CONTENTS ...
Page 17: ...PN 4237395AA 1 6 USE AND FUNCTION MATERIAL SAFETY DATA SHEETS MSDS ...
Page 43: ...PN 4237395AA 3 4 OPERATION PRINCIPLES TYPES OF PRIMING ...
Page 47: ...PN 4237395AA 4 4 SPECIFICATIONS CHARACTERISTICS PERFORMANCE CHARACTERISTICS ...
Page 49: ...PN 4237395AA A 2 LOG SHEETS ...
Page 51: ...PN 4237395AA A 4 LOG SHEETS ...
Page 53: ...PN 4237395AA A 6 LOG SHEETS ...
Page 55: ...PN 4237395AA A 8 LOG SHEETS ...
Page 69: ...PN 4237395AA B 14 OPERATOR CERTIFICATION PROCEDURE TQ PREP WORKSTATION PRECISION PROCEDURE ...
Page 71: ...PN 4237395AA B 16 OPERATOR CERTIFICATION PROCEDURE TQ PREP WORKSTATION PRECISION PROCEDURE ...
Page 73: ...PN 4237395AA 2 REFERENCES ...
Page 77: ...PN 4237395AA 4 GLOSSARY ...
Page 85: ......