![Beckman Coulter TQ-Prep Reference Download Page 13](http://html.mh-extra.com/html/beckman-coulter/tq-prep/tq-prep_reference_2719765013.webp)
PN 4237395AA
1-2
USE AND FUNCTION
SYSTEM COMPONENTS
1.3
SYSTEM COMPONENTS
The system has three main work areas:
r
Sample processing area
r
Reagent storage and dispense area
r
Touch screen.
Sample Processing Area
See Figure 1.2. The sample processing area consists of a 32-tube capacity removable carousel,
an indexing base, dispensing head with tube detector, sensing and sequencing devices, a tube
lifter/vortex mixer, a reagent delivery system and electronics. The lid to the sample processing
area has a safety interlock switch so sample processing can only occur with the lid closed.
Figure 1.2 Sample Processing Area
Carousel
Two sample carousels, each with a handle and 32-test tube slots, are provided with each unit.
The key located on the carousel aligns with the hub key on the indexing base to correctly
position the carousel for sample processing.
Indexing Base
The indexing base moves the carousel to its home position and moves the carousel during
sequential sample processing.
Dispensing Head
The retractable dispensing head adds a precise amount of multiple reagents to the sample.
7395020A
CAROUSEL
HUB KEY
DISPENSING HEAD/
TUBE DETECTOR
TUBE LIFTER/
VORTEX MIXER
INDEXING
BASE
KEY
This label is to
notify you that a
biohazardous
condition can exist
in this area if
blood is spilled
SAMPLE
PROCESSING
AREA LID
Summary of Contents for TQ-Prep
Page 4: ...PN 4237395AA iv REVISION STATUS ...
Page 8: ...PN 4237395AA viii CONTENTS ...
Page 17: ...PN 4237395AA 1 6 USE AND FUNCTION MATERIAL SAFETY DATA SHEETS MSDS ...
Page 43: ...PN 4237395AA 3 4 OPERATION PRINCIPLES TYPES OF PRIMING ...
Page 47: ...PN 4237395AA 4 4 SPECIFICATIONS CHARACTERISTICS PERFORMANCE CHARACTERISTICS ...
Page 49: ...PN 4237395AA A 2 LOG SHEETS ...
Page 51: ...PN 4237395AA A 4 LOG SHEETS ...
Page 53: ...PN 4237395AA A 6 LOG SHEETS ...
Page 55: ...PN 4237395AA A 8 LOG SHEETS ...
Page 69: ...PN 4237395AA B 14 OPERATOR CERTIFICATION PROCEDURE TQ PREP WORKSTATION PRECISION PROCEDURE ...
Page 71: ...PN 4237395AA B 16 OPERATOR CERTIFICATION PROCEDURE TQ PREP WORKSTATION PRECISION PROCEDURE ...
Page 73: ...PN 4237395AA 2 REFERENCES ...
Page 77: ...PN 4237395AA 4 GLOSSARY ...
Page 85: ......