82 MC6-Ex User Manual
– Part 9, Additional Information
Calibration ranges, corresponding coefficients and examples of calibration
points:
Calibration
Corresponding
Ranges
(
*
Coefficients Example of fixed calibration points
(#
, °C
Negative (sub-)ranges:
-189 … 0 °C
a
4
, b
4
–189.3442, –38.8344,
0.01
-38 … 30 °C
a
5
, b
5
(
†
–38.8344,
0.01,
29.7666
Positive (sub-)ranges:
0 … 30 °C
a
11
0.01,
29.7666
-38 … 30 °C
a
5
, b
5
(
*
–38.8344,
0.01,
29.7666
0 … 157 °C
a
10
0.01,
29.7666,
156.5985
0 … 232 °C
a
9
, b
9
0.01,
156.5985,
231.928
0 … 420 °C
a
8
, b
8
0.01,
231.928,
419.527
0 … 660 °C
a
7
, b
7
, c
7
0.01,
231.928,
419.527, 660.323
0 … 962 °C
a
6
, b
6
, c
6
, d
0.01,
231.928,
419.527, 660.323, 961.78
†)
sub-range 5 is presented twice since it needs to be entered separately both for negative and
positive sides.
Note.
If your calibration certificate include two sets of coefficients. One
for zero current and another for 1 mA current, enter the latter
ones into MC6-Ex.
*
)
The range limits in the adjacent list are rounded and
shown as they are on MC6-Ex User Interface.
#)
Not all calibration laboratories necessarily use the
same points. These are for reference only.
When selecting
ITS 90 PRT (R
tpw,…)
as the
Sensor Calculation Formula
, the
Sensor Range
entered in the 1
st
configuration page defines how many additional
User Sensor
pages is added to MC6-Ex. If the range includes temperatures be-
low zero, the total amount of configuration pages is four:
1. First page for general settings. See page 80.
2. Second page for entering the
R
tpw
constant value.
3. Third page for selecting the deviation formula for
negative
temperature
sub-range and entering the coefficients (a
4
and b
4
or a
5
and b
5
).
4. Fourth page for selecting the deviation formula for
positive
temperature
sub-range and entering the coefficients (a
5
, a
6
, a
7
… etc.).
When no negative sub-range is included in the calibration, the third page men-
tioned above is not included and the total amount of configuration pages is three.
Examples of coefficients a
7
, b
7
and c
7
.
Summary of Contents for MC6-Ex
Page 2: ......
Page 37: ...Things discussed in this part Presenting the Meter and how to take it into use Part 3 METER ...
Page 60: ...52 MC6 Ex User Manual Part 5 Documenting Calibrator ...
Page 96: ...88 MC6 Ex User Manual Part 9 Additional Information ...