52
6-847718-00 Rev. B00
Wiring Schematic 7: Multi-Signal Module Relay Outputs
SIGNAL 1
SIGNAL 2
SIGNAL 3
SIGNAL 4
SIGNAL 5
Dry normally-closed
contacts at source
equipment or line
pressure switches
COM
+
COM
+
COM
+
COM
+
COM
+
Multi-Signal Module
with relays
20 wire
ribbon cable
to adjacent
module
SIGNAL 1 - pins 1 & 2
SIGNAL 2 - pins 3 & 4
SIGNAL 3 - pins 5 & 6
SIGNAL 4 - pins 7 & 8
SIGNAL 5 - pins 9 & 10
Dry normally-closed
output relay contacts
NOTE:
Output relay contacts are dry normally closed
type.
Contacts open upon designated alarm activation.
Contacts remain open until alarm condition is cor-
rected.
Pressing MUTE
on Annunciator Module WILL
deactivate relay until audible alarm is again reacti-
vated.
Relay contact ratings are 2 A @ 30 VDC/0.5 A @ 125
VAC.
NOTE:
Field wiring cable shields must be grounded at
only one end, inside alarm panel back box. Refer
to page 39 for details.
W
iring
Schematics
Summary of Contents for TotalAlert 2
Page 1: ...Installation Operation and Maintenance Instructions 6 847718 00 Rev B00...
Page 2: ......
Page 22: ...20 6 847718 00 Rev B00...
Page 97: ...95 6 847718 00 Rev B00 Figure 86 Clear Network Accepted Operation...
Page 99: ...97 6 847718 00 Rev B00 Operation Figure 89 Event Log Cleared...
Page 101: ...99 6 847718 00 Rev B00 Operation Figure 91 Transfer Setup Figure 92 Transfer Setup Complete...
Page 129: ...127 6 847718 00 Rev B00 Troubleshooting Guide Troubleshooting...
Page 130: ...128 6 847718 00 Rev B00...
Page 139: ...137 6 847718 00 Rev B00 Replacement Parts Notes...
Page 143: ......
Page 144: ...2007 BeaconMed s All rights reserved 6 847718 00 Rev B00...