
117
6-847718-00 Rev. B00
Press and hold the front panel TEST but-
ton for 5 s to initiate the following
sequence:
•
Audible alarm sounds for 1 s as
TT EE SS TT
is
displayed.
•
LED display segments and LOW,
NORMAL, and HIGH indicators illuminate
in sequence for 0.5 s.
•
Firmware revision (e.g.
2
2.. 0
0.. 0
0
) is displayed
for 2 s.
•
A
A D
D D
D RR
is displayed for 1 s then module’s
network address is displayed for 2 s.
•
Units of measure setting (e.g.
P
P SS II
) is
displayed for 2 s.
•
LL OO
is displayed for 1 s, LOW alarm LED is
illuminated, and current low alarm set point
is displayed for 2 s.
•
K
K ,,
is displayed for 1 s, HIGH alarm LED is
illuminated, and current high alarm set
point is displayed for 2 s.
•
Module returns to monitoring mode.
Digital Display Module
Summary of Contents for TotalAlert 2
Page 1: ...Installation Operation and Maintenance Instructions 6 847718 00 Rev B00...
Page 2: ......
Page 22: ...20 6 847718 00 Rev B00...
Page 97: ...95 6 847718 00 Rev B00 Figure 86 Clear Network Accepted Operation...
Page 99: ...97 6 847718 00 Rev B00 Operation Figure 89 Event Log Cleared...
Page 101: ...99 6 847718 00 Rev B00 Operation Figure 91 Transfer Setup Figure 92 Transfer Setup Complete...
Page 129: ...127 6 847718 00 Rev B00 Troubleshooting Guide Troubleshooting...
Page 130: ...128 6 847718 00 Rev B00...
Page 139: ...137 6 847718 00 Rev B00 Replacement Parts Notes...
Page 143: ......
Page 144: ...2007 BeaconMed s All rights reserved 6 847718 00 Rev B00...