Wiring (Cont.)
Relay Board Outputs
Inst
allation
Procedures
A master or combo alarm with 32 signals
has only one relay board. Signals are
numbered A1 through A32.
A master or combo alarm with 64 signals
has two relay boards. Signals are labeled
A1 through A32 on relay board A and B1
through B32 on relay board B.
1. Identify each pair of field installed relay
output signal wires inside alarm panel
back box assembly.
2. Route each pair of relay output signal
wires as shown in Figure 25 to
appropriate terminals on relay board(s).
3. Connect each pair of signal wires to
desired relay board terminals. The
terminals are labeled “1” through “32”.
Refer to
Wiring Schematic 4
(Page 49).
Figure 25: Relay Board Wire Routing