Troubleshooting
Symptom
Possible Cause
Corrective Action
24)High and low alarm with
-- FF 44 --
display on digital display
module.
Gas/vacuum service mismatch
between digital display module
and digital sensor module as a
result of:
a. Sensor wired to wrong digital
display module.
b. Digital display module
programmed with the wrong
service.
a. Correct sensor wiring.
b. Program digital display module
for the correct gas/vacuum
service. Refer to
Set-Up
instructions (page 67).
25)High alarm with
11 99 00
to
22 00 00
display on digital display
module.
a. Digital sensor connected to
digital display module with gas
ID set to
N O N E.
a. Program gas ID in digital
display module.
26)No blinking LED on digital
sensor module.
No power to digital sensor due
to:
a. Faulty wiring.
b. Faulty digital sensor module.
a. Correct sensor wiring.
b. Replace digital sensor module.
27)Fast blinking LED on digital
sensor module.
Sensor out of range:
a. Vacuum digital sensor module
connected to pressure service.
b. Pressure digital sensor module
connected to vacuum service.
c. Faulty digital sensor module.
a. Verify piping connections.
Vacuum sensor maximum
overpressure: 30 psi.
b. Verify piping connections.
c. Replace digital sensor module.
28)VFD indicates
where
DDDDDDDDDDDD
indicates the device name
(or serial number if device
name is unknown)
Lost connection with
Device: DDDDDDDDDDDD
a. A remote alarm panel has
become disconnected from the
Ethernet network.
b. A remote alarm panel has been
intentionaly removed or
replaced.
a. Reconnect remote alarm to the
network.
b. Perform a Clear Network to
force alarm panel to re-scan for
remote devices. Refer to Set-
Up instructions:
Page 59 - Front panel controls
Page 94 - Web pages
29.VFD stalled with
Booting...
a. Annunciator module firmware
corrupted (possibly due to failed
firmware update attempt).
b. Faulty annunciator module.
a. Reprogram module firmware
(BeaconMedaes personnel
must perform this procedure).
b. Replace annunciator module.
Troubleshooting Guide (Cont.)
Summary of Contents for TotalAlert 2
Page 1: ...Installation Operation and Maintenance Instructions 6 847718 00 Rev B00...
Page 2: ......
Page 22: ...20 6 847718 00 Rev B00...
Page 97: ...95 6 847718 00 Rev B00 Figure 86 Clear Network Accepted Operation...
Page 99: ...97 6 847718 00 Rev B00 Operation Figure 89 Event Log Cleared...
Page 101: ...99 6 847718 00 Rev B00 Operation Figure 91 Transfer Setup Figure 92 Transfer Setup Complete...
Page 129: ...127 6 847718 00 Rev B00 Troubleshooting Guide Troubleshooting...
Page 130: ...128 6 847718 00 Rev B00...
Page 139: ...137 6 847718 00 Rev B00 Replacement Parts Notes...
Page 143: ......
Page 144: ...2007 BeaconMed s All rights reserved 6 847718 00 Rev B00...