3
Installation instruction
3.1
Introduction
Install the exhauster unit in an area where the noise levels do not cause an inconvenience and adequate
ventilation is available for cooling purposes. The Anaesthetic Gas Scavenging disposal system should be floor
mounted. Bolts used for mounting should be M 8 or similar size.
Before commencing installation, check if the electrical data on the pump specification plate/s
(positioned under the electrical connection box) are compatible with the local power supply.
Before electrically connecting the AGS plant, ensure that the power supply is off and correctly
isolated.
The electrical power supplies to the AGS plant must be connected by a qualified electrician in
accordance with the wiring diagram supplied with the plant. All wiring must be in accordance
with IEE regulations. Cable sizes given in section
Electric data
are recommendations only.
The (electric components of the) device should be installed outside the patient environment.
This environment is defined as a hemisphere with a radius of 2 meter around the patient.
Installation proposal AGS Simplex
Instruction book
22