29
Medical Gas Central Alarm - Medipoint 125 Digital Alarm
4233500107.03
TABLE 6: MEDIPOINT 125 DIGITAL - ALARM LOGIC
Condition
Alarm panel displays
Normal
NORMAL
Power LED
Green, On
System Fault
Black, Off
Gas Status Bar Green, Steady
Audible
Off
Alarm
Condition
Gas service fault (flashing)
Example shows gas service fault 1.
For full list of standard gas service faults see
table 7.
Gas service fault 2 and 3 are also yellow
flashing.
Power LED
Green, On
System Fault
Black, Off
Gas Status Bar Red, Flashing
Audible
Active
Pressure fault (flashing)
Example shows gas service fault 4.
Power LED
Green, On
System Fault
Black, Off
Gas Status Bar Red, Flashing
Audible
Active
Audible
Mute
Gas service fault (flashing) Mute activated
Audible
Silenced for 15 minutes
System
fault
Gas service Line contact fault.
Gas service fault (steady)
SYSTEM ALARM (flashing)
Audible
Example shows gas service fault 1.
Power LED
Green, On
System Fault
Red, Flashing
Gas Status Bar Red, Steady
Audible
Active
Press TEST switch
All alarm conditions will flash, the condition with
the line contact fault will flash slower than the
other gas status bars, and audible will sound.
Alarm conditions
Test switch
Test button
Held
Power
failure to
panel
Power LED
Black, Off
System Fault
Red, Flashing
Gas Status Bar Blank
Audible
Active