Medical Gas Central Alarm - Medipoint 125 Digital Alarm

4233500107.03

5. 

Operation of the Medipoint 125 

Digital alarm.

5.1 

User operation.

The following section details instructions for use once the 

alarm has been configured and fully operational.  Figure 

40 details the user interface for the main operating touch 

screen and fascia strip that contains LED indicators and 

push buttons.

Figure  40  -  Example,  6  gas  setup.  User  interface 

shown in normal operating condition.

Fascia strip - Characters/Icons:

Icon

Description

Power

LED

Green, illuminated - Power On

Black, de-energised - Power Off

Mute 

Button

Press to activate to silence the 

audible alarm.

System

Fault

Black, de-energised - System Normal

Red, illuminated - System Fault

Test 

Button

Press to activate the alarm test 

button.

Touch Screen - Indication/Warnings:

Icon

Description

Gas service legend, 

colour ID bar and 

location served
Normal status

Gas service fault 1 status
Gas service fault 2 status
Gas service fault 3 status
Gas service fault 4 status

Backup 

Battery

Backup battery Status, 

charging.

Backup battery Status, 

fault.

Date / 

Time

Date time not setup.  

Only visible if not set, 

see section 3.

The  green  NORMAL  status  bar  is  a  steady  light,  which 

changes  to  a  red  flashing  bar  during  HIGH  and  LOW 

PRESSURE  faults  to  draw  attention  to  the  failure.  

Pressure  fault  conditions  are  accompanied  by  a  two 

tone audible warning.  See table 6.

The  audible  warning  continues  until  either  the  MUTE 

switch  is  operated  or  the  gas  pressure  returns  to 

NORMAL. If the MUTE facility is operated and the alarm 

condition  remains,  the  audible  warning  will  re-sound 

after  15  mins  and  a  further  MUTE  selection  will  be 

required.  Should  a  further  alarm  condition  occur  after 

the panel has indicated a fault and has been muted, the 

audible warning is re-Activated.

Operation of the TEST switch changes all status bars to 

flashing and sounds the audible two tone warning.

 

When  the  slave  relay  connection  is  interconnected  to 

either a Medipoint Digital central alarm system or other 

suitable  system,  the  contacts  are  closed  when  all  gas 

services  are  indicating  NORMAL.  In  the  event  of  any 

alarm condition, the relay contacts open, breaking the 

circuit to the remote system. The relay contacts remain 

open  until  all  alarm  conditions  return  to  the  NORMAL 

status.  There  is  also  a  line  contact  monitor  integral  to 

the board available across the N/O and C contacts. This 

function is selected within the on screen setup menus 

see section 3.  

The green POWER ON LED is normally illuminated and is 

extinguished in the event of an electrical power failure. 

The red SYSTEM ALARM LED is normally extinguished and 

illuminates flashing together with an audible warning in 

the event of any of the following failures:-

•  Alarm contact line fault.

•  Electrical power supply failure. 

•  Internal alarm circuit failure.

When  a  line  contact  fault  is  detected,  the  NORMAL 

status bar changes to red (steady).  The SYSTEM ALARM 

LED  illuminates  (flashing)  accompanied  by  an  audible 

warning. 

Operation of the TEST switch changes all gas status bars 

to  Red  (flashing).    If  there  is  a  faulty  sensor  or  wiring 

present  the  flashing  rate  will  be  at  half  speed.    This 

facility is designed to aid fault diagnosis. The Medipoint 

125 medical gas alarm logic is detailed in Table 6.

Summary of Contents for 8102371401

Page 1: ...Operation and Maintenance Instructions Medical Gas Central Alarm Medipoint 125 Digital Alarm Part number 4233500107 Revision 03 Jan 2021...

Page 2: ...hone 44 0 1246 474242 Email gbn info beaconmedaes com Website Contacts www beaconmedaes com BeaconMed s reserves the right to make changes and improvements to update products sold previously without n...

Page 3: ...in screen 3 6 Exit Setup Screen 3 7 Confi guration Menu Map 4 0 Commissioning 4 1 Alarm panel test 4 2 Checking the slave relay output 4 3 Checking the remote audible warning 4 4 Medical gas alarm fau...

Page 4: ...ckaged and stored in controlled conditions Environmental Operating Conditions Adverse environmental conditions and harsh abrasives or chemicals may cause damage to the unit Environmental Protection Di...

Page 5: ...rammed from on screen menus to provide a fully flexible alarm system Allowing the panels to be adapted to suit any hospital s medical gas monitoring requirements see section 1 12 for programming detai...

Page 6: ...nt of any defect within the monitored functions the SYSTEM ALARM LED illuminates flashing accompanied by the audible warning and the defective gas service alarm display status bar illuminates An inter...

Page 7: ...operty and shall not without our permission be altered ATLAS COPCO STANDARD Class Tolerances if not indicated according to Material Treatment Name Scale Family Drawing owner Secrecy Class 1102 K Compa...

Page 8: ...p to four fault conditions yellow or red flashing alarm conditions to show the gas service status All alarm conditions are illuminated by a flashing display and accompanied by an audible warning Each...

Page 9: ...on these PCB s which are interconnected by means of a multi way ribbon cable and polarised connector see figure 2 3 The area central alarm internal electrical installation complies with all relevant...

Page 10: ...cular displayed alarm condition only This facility is designed for use when the plant or pipeline is shut off for a prolonged period On returning the pipeline pressure to NORMAL this feature automatic...

Page 11: ...nnections have 5 terminals each See table 2 for details for standard options Each gas service and fault legend is also fully customisable through the on screen setup menus TABLE 2 GAS SERVICE CONNECTI...

Page 12: ...uired by the contract specification Feed cables into the box leaving 400mm to 75 150 105 105 7 260 5 TOP 40 40 40 40 17 20 KNOCKOUT SEGMENTS 200 17 45 2 0 KNOCKOUT SEGMENTS BACKBOX SURFACE Figure 10 B...

Page 13: ...Power supply PCB installation No Description QTY 1 Back Box 1st fix supply N A 2 Power supply PCB 2nd fix supplied loose 1 3 PCB Spacer 2nd fix supplied loose 4 4 M3 Nut 2nd fix supplied loose 4 5 M3...

Page 14: ...nections No Description QTY 1 Earth cable 2nd fix supplied assembled 1 2 M3xNut 2nd fix supplied loose 4 3 M3 washer 2nd fix supplied loose 4 Typical earth stud assembly 2 2 3 1 Alarm Door Earth Stud...

Page 15: ...igure 15 Alarm data connection MP125D 1 S1 ON IN OUT A B A B S MP125D 2 S1 OFF IN OUT A B A B S MP125D 3 S1 ON IN OUT A B A B S Note S terminal is for connecting the screen cable 2 2 6 Remote audible...

Page 16: ...mmissionedimmediately disconnect the standby battery Connect electrical power supply wires to electrical power supply plug terminals E N and L 1 2 Figure 21 Alarm door security fastener No Description...

Page 17: ...aracters Icons Icon Description Enter Key TAB key Moves to next field and highlights text Highlighted text will be overwritten on next key press Tick key Apply update and takes you to the main screen...

Page 18: ...e Alarm setup 2 screen or press the Exit Icon to go back to the main operating screen 3 4 Alarm Setup Screen 1 Figure 26 The following steps describes the setup process for the alarm setup screen 1 3...

Page 19: ...e new numerical password Press the TAB key or press the next field to go to the next field Confirm Use the num pad to enter the password again Press the ENTER key or press anywhere on the screen away...

Page 20: ...setup process for the main gas service setup screen Figure 27 Main Gas Service Setup screen Special Characters Icons Icon Description Previous Icon Takes you to the previous menus screen Next Icon Tak...

Page 21: ...k to set as inputs tick to set as remote transmission Press to toggle between the options 3 5 6 Location Special Characters Icons Description Exit Icon Takes you back to the main screen O2 Oxygen Gas...

Page 22: ...elds will be editable 3 5 7 1 Standard gas options These fields allow for selecting the source equipment type available for the standard gas type selected Press on the matching equipment type to auto...

Page 23: ...etup screen Figure 33 Main Gas Service Setup screen Special Characters Icons Icon Description Previous Icon Takes you to the previous menus screen Next Icon Takes you to the next menus screen Exit Ico...

Page 24: ...key press Previous Icon Takes you to the previous menus screen Next Icon Takes you to the next menus screen Exit Icon Takes you back to the main screen Reject Icon Rejects changes Accept Icon Accepts...

Page 25: ...6 shows examples of all status flashing See table 6 in section 5 Operation of the Medipoint 125 Digital alarm for troubleshooting Figure 36 Alarm Test Button example 6 gas 4 2 Checking the slave relay...

Page 26: ...e table 2 section 1 8 for full list of standard gas service fault legends Check that the MUTE switch operates satisfactorily Press the audible alarm will be silenced and the mute icon will show on scr...

Page 27: ...ED is flashing red The audible alarm is active See figure 39 fro example Figure 39 Example Gas service 1 connection fault 4 5 3 Input Reconnect Reconnect the input plug to gas channel System will retu...

Page 28: ...d and the alarm condition remains the audible warning will re sound after 15 mins and a further MUTE selection will be required Should a further alarm condition occur after the panel has indicated a f...

Page 29: ...mple shows gas service fault 4 Power LED Green On System Fault Black Off Gas Status Bar Red Flashing Audible Active Audible Mute Gas service fault flashing Mute activated Audible Silenced for 15 minut...

Page 30: ...lank normal Yellow flashing fault Fault status 3 Blank Fault status 4 Blank normal Red flashing fault Manifold Control Systems Normal status Green normal blank during fault Fault status 1 Blank normal...

Page 31: ...to the main screen automatically if a new alarm condition becomes active or after one minute if the alarm is not manually exited from the information screen 5 2 2 Data Logging Screen The data logging...

Page 32: ...s distributors will be able to supply similar contracts in other areas CAUTION It is essential that only genuine BeaconMed ssparepartsarefittedduringmaintenance WARNING Obtain a work permit or equival...

Page 33: ...33 Medical Gas Central Alarm Medipoint 125 Digital Alarm 4233500107 03 Figure 45a Alarm panel assembly details...

Page 34: ...34 Medical Gas Central Alarm Medipoint 125 Digital Alarm 4233500107 03 Figure 45b Alarm panel assembly details...

Page 35: ...e 5 year maintenance replace the LCD screen and backup butteries See maintenance procedures in 6 1 and figures 45a 45b for details 8 RECOMMENDED SPARES Table 8 details the list of recommended spares t...

Page 36: ...2460 Atlas Copco Medical Ltd Telford Crescent Staveley Derbyshire S43 3PF UK Tel 44 0 1246 474242 www beaconmedaes com Atlas Copco Airpower n v Boomsesteenweg 957 2610 Wilrijk Belgium...

Reviews: