Medical Gas Central Alarm - Medipoint 125 Digital Alarm
4233500107.03
Safety Precautions
WARNING! DO NOT USE OIL OR GREASE on
any parts in contact with the medical gases for any
reason. This could lead to a FIRE or an EXPLOSION.
Only use approved OXYGEN COMPATIBLE lubricants,
which can be purchased from BeaconMedæs if
necessary.
Operator should have carefully read and become familiar
with the contents of this manual before maintaining the
Medipoint 125 Digital Alarm.
Operator is expected to use common sense safety
precautions, good workmanship practices and follow
any related local safety precautions.
Component descriptions and parts lists are available on
request.
Identification of symbols
The following symbols apply to this product and are used
in these instructions and on the product in question. The
meanings of these symbols are as specified below: -
Read instructions
Ambient temperature range
Ambient humidity range
Ambient pressure range
Date of manufacture
Do not dispose of in general waste
Electromagnetic Interference
The panel has been tested to BS EN 61326-1-
Electromagnetic compatibility - Requirements and tests.
Environmental Transport and Storage
Conditions
All products are separately packaged and stored in
controlled conditions.
Environmental Operating Conditions
Adverse environmental conditions and harsh abrasives
or chemicals may cause damage to the unit.
Environmental Protection
Discard the unit and/or components in any standard
refuse facility. The unit does not contain and hazardous
substances.
Cleaning
The alarm cover should be wiped over with a damp cloth
frequently to remove any dust or foreign substances.
Electrical Details
CAUTION! It is necessary to check the integrity of
the power source for safety at regular intervals. These
checks should be carried out annually and replacement
power supplies used as necessary.
Power source
Mains operated using 110V-230V, 50/60Hz, alternating
current.
Cable size:
Frequency Voltage
Minimum Cable Size
50 Hz
230 V
3 x 1.5 mm
2
60 Hz
110 V
3 3 AWG14
Current requirements - 3.0 amps
Type of protection against electric shock.
Class 1 (Mains supplied equipment using a protected
earth).
Relative Humidity 90% max.
Altitude up to 2000m
Pollusion Degree 2
Mode of operation:
• Indoor use.
• Continuous (equipment may be left switched on
indefinitely).
Note - A double pole fused isolation switch or circuit
breaker must be installed locally to the medical gas
alarm. The isolation point must be clearly marked for
the equipment it is intended for, and within easy reach.
The switch must require a tool or key to operate to
prevent unauthorized isolation of the device.