BD FACSFlow Supply System User’s Guide
40
Sheath filter
components
Removing the old
filter
To remove the old filter:
1. Place the cytometer in standby mode.
2. Remove the sheath filter assembly by pressing the
quick-disconnect on both sides of the filter assembly, and
pulling it out of the holder on the sheath plenum enclosure.
3. Over a sink or beaker:
• Remove the vent line from the filter and set it aside.
• Remove the filter base and set it aside.
4. Discard the used filter assembly in an appropriate receptacle.
Item
Description
1
Filter base
2
Filter
3
Quick-disconnect
4
Fluid line
5
Vent line
6
Vent cap
1
2
3
4
5
6