BD Alaris™ neXus GP Volumetric Pump
Advanced Features
BDDF00558 Issue 5
36/72
Profile Filter
This option allows selected profiles to be filtered from view . It can be enabled/disabled in the data set via the BD Alaris™
neXus Editor .
1 . Press the
d
button to access the
OPTIONS
menu .
2 . Select
PROFILE FILTER
using the
f
keys, press the
OK
softkey to confirm .
3 . Select profile that requires filtering using the
f
keys .
4 . Press
MODIFY
to disable profile and press
OK
to confirm or
QUIT
to return to main display .
Note:
If only one profile is available and all others are disabled in this option, the
CONFIRM PROFILE
screen will not
display upon start up .
Standby
This option allows the Pump to be placed on standby mode . It can be enabled/disabled in the data set via the BD Alaris™
neXus Editor .
1 . Press the
d
button to access the
OPTIONS
menu .
2 . Select
STANDBY
using the
f
keys, press the
OK
softkey to confirm .
3 . Select
CANCEL
to return to main display .
Date and Time
This option shows the Pump’s current date and time . This can be updated via Pump’s Service mode by Qualified Service
Personnel, or will automatically update when the Pump connects to the BD Alaris™ Communication Engine, if one is
deployed at the facility .
Network Status
This option allows information about the wireless network connection status to be viewed by Qualified Service Personnel .
Refer to the
Technical Service Manual
for the description of network data displayed on these screens .
1 . Press the
d
button to access the
OPTIONS
menu .
2 . Select Network Status using the
f
keys . Press
OK
to confirm .
3 . Enter password to see the
NETWORK STATUS
information .
4 . Navigate through the pages using the softkeys .
5 . Select
QUIT
to return to the main display .
Access code to display NETWORK STATUS menus should only be entered by Qualified Service Personnel.
Note:
For information on Network Status codes, please refer to the
Technical Service Manual
.
Wireless Connection
The Pump disables wireless communication when docked to a Workstation, and will automatically attempt to re-connect
if enabled, and removed from the Workstation . This option allows for wireless communication to be enabled/disabled
manually .
Note:
It may take up to five minutes for the Pump to reconnect to a Wi-Fi network if connection was disabled manually .
Note:
If Wi-Fi is disabled manually, that setting will remain through power cycles .
1 . Press the
d
button to access the
OPTIONS
menu .
2 . Select
WIRELESS CONNECTION
using the
f
keys . Press
OK
to confirm .
3 . To toggle the
WIRELESS CONNECTION
status between enabled and disabled press
MODIFY
.
4 . Press
OK
to confirm . Press
QUIT
to disregard changes and return to the main display .