Display
Action
AIR IN LINE
Single air bubble exceeds alarm limit.
Set not fitted correctly into air in line
detector.
• Ensure set is fitted correctly in the air-in-
line detector.
• Inspect infusion set for air bubbles and
take appropriate action.
• Remove air according to hospital policy.
• Check fluid level in container.
• Check fluid level in drip chamber.
• Restart infusion.
AIR IN LINE
Accumulated air bubbles exceeds
alarm limit.
• Inspect infusion set for air bubbles and
take appropriate action.
• Remove air according to hospital policy.
• Check fluid level in container.
• Check fluid level in drip chamber.
• Restart infusion.
DOOR OPEN
Door was opened during an infusion.
• Close door or clamp infusion set using
roller clamp.
• Restart infusion.
DOWNSTREAM OCCLUSION
A blockage has occurred downstream.
• Check fluid path between Pump and
patient for clamps, connectors, kinks or
blockages.
• Examine access site for signs of
complications e.g. redness, swelling, pain,
heat.
NO FLOW
Flow sensor detects no flow.
• Check flow sensor.
• Check fluid level in container.
• Ensure all clamps above Pump are open.
• Check fluid level in drip chamber.
• Ensure that the bag spike is inserted
correctly.
• Check flow sensor is clean.
DO NOT USE
Internal error.
• Remove Pump from use.