Action
UPSTREAM OCCLUSION
A blockage has occurred upstream.
Possible container empty.
• Check that the fluid container is at a
minimum height of 300mm above the
pump.
• Ensure the bag spike is fully inserted into
the fluid container and not obstructed.
• Ensure the fluid container is not empty.
• Ensure air vent on drip chamber is open
on all glass and semi rigid containers.
• Check there are no kinks in the infusion
sets above the Pump.
• Check all infusion set clamps above the
Pump are fully open.
• Check fluid level in the drip chamber.
• Open the Pump door and ensure the
infusion set is loaded correctly.
• Consider if the 15µm filter in the drip
chamber is blocked or partially blocked.
• Ensure the infusion set has not been used
for more than 72 hours.
• If alarm continues consider changing the
infusion set.
FLOW ERROR
Gross difference between detected
drops and expected amount of drops.
• Clamp infusion set using roller clamp.
• Check flow sensor.
• Check fluid level in drip chamber.
FLOW ERROR
(In Secondary infusion mode only)
Unexpected drops detected.
• Hang secondary container above primary.
• Check drops are from secondary container
when infusing.
• Flow sensor disconnection is
recommended.
FREE FLOW
Uncontrolled flow possible.
• Clamp infusion set using roller clamp.
• Remove Pump from use.
Alarms c
ontinued