0000CF04424 Iss. 1 11/18
Note:
XX is the
AUTO OFFSET
pressure and is determined by the user. This
adjustment,
AUTO OFFSET
value 15-100mmHg, is configurable
by profile within the data set. At pressures up to 100mmHg the
AUTO OFFSET
value is added. For pressures above 100mmHg the
downstream occlusion alarm limit is set to whatever the
AUTO
OFFSET
value is as a percentage above the average infusion line
pressure up to the maximum occlusion alarm limit
defined within the
profile
.
Pressure Trend
TREND
71
mmHg
–
+
BACK
17:15
0
1. To check the pressure trend graph press the
e
button.
2. Press the
TREND
softkey to view the pressure trend of
the previous 12.5 hours of clock time. The pressure trend
can be viewed at 15 minute intervals by using the
+/-
softkeys. The pressure trend is updated every 15 minutes
and can be viewed with greater resolution. The pressure
trend graph displays the line pressure at a given time.
3. Press the
OK
softkey to exit the pressure screen.