IDS701 User Manual
Page 12 of 25
4.2. Step of calibration operation:
According to the second method which can enter setting menu,
C01-C39
step
Method of operation
display
Remark
1
[C01 ]
After you enter to calibration
mode, it display [C01 ]
2
press
[C1 1]
Weight unit
option
:
1=kg
2=lb
3
press
press
press
or
[C02 ]
[C02 0]
[C02 2]
Set decimal digits
option
:
0/1/2/3/4
Select decimal digit
example
:
two decimal point
:
[C02 2]
4
press
press
press
or
[C03 ]
[C03 1]
[C03 5]
Set graduation
option
:
1/2/5/10/20/50
Select required graduation
example
:
graduation 5
:
[C03 5]
5
press
press
press
or
/
[C04 ]
[0100.00]
[0100.00]
Max capacity
example
:
max weighing 100kg:
[0100.00]
6
press
press
press
press
[C05 ]
[C05 0]
[C05 1]
[CAL 9]
。
。。
。
。
。
Zero calibration
option
:
0=non-calibration
zero
1=need calibration zero
calibration zero please choose 1 and
ensure scale is empty and “stable”
Manual usuario BBG ref BULL