Operation Manual
8 - 37
8
Powering Up and Shutting Down the System
Q
Q
Q
"Restore from Injector Shutdown"
Q
"Restore from System Shutdown"
8.1 Powering Up the System
1.
Press the power button on the workstation. The LED turns green and the display powers up.
2.
Ensure the rocker switch on the Scan Room Unit Power
Supply is set to ON (“I” is pressed in). The injector powers up.
3.
The display will briefly show the system communication
statuses and software versions.
NOTE:
If a communication error is displayed, contact
Bayer.
4.
Next, the Safety Screen will display. Read the warnings on
the Safety Screen and select
Continue
.
5.
The system powers up and the Home Screen is displayed.
Figure 8 - 1: Safety Screen
CAUTION
Electric Shock Hazard - Minor or moderate patient and/or worker injury may result.
Q
Verify that the voltage and frequency as labeled matches the voltage and frequency of the electrical outlet.
Summary of Contents for MEDRAD MRXperion