Operation Manual
3 - 9
3
Warnings, Cautions, and Notices
Q
Q
Q
3.1 Warnings
WARNINGS
Air Embolism Hazard - Serious patient injury or death may result.
Q
Use only disposables supplied by Bayer. Refer to
"Section 19.2 - MEDRAD® MRXperion MR Injection System Syringe Kits"
.
Q
Use only catheters and connectors with pressure ratings that are compatible with this system. Refer to
"Section 18.5 - System Capabilities"
for programmable pressure limit options.
Q
Use only accessories and options approved by Bayer.
Q
The operator needs to check for air to confirm that the syringes and tubing have been inspected for the presence of air.
Environmental Contamination Hazard - Serious patient or worker injury or death may result.
Q
Visually inspect all components before use.
Q
Do not use damaged components.
Electric Shock Hazard - Serious patient and/or worker injury or death may result.
Q
Only use the power cord supplied with the system or by the Bayer representative.
Q
Equipment must only be connected to a supply mains with protective earth.
Q
Do not remove or open any enclosure, as hazardous voltages exist within system components.
Q
Do not immerse any system components in water or cleaning solutions.
Q
Disconnect the system from supply mains before cleaning.
Q
Do not use the system if any worn or damaged cabling is detected. Contact Bayer for replacement.
Fire Hazard - Serious patient and/or worker injury or death may result.
Q
Use the correct fuse type, replaced by Bayer or personnel trained by Bayer.
Explosion Hazard - Serious patient and/or worker injury or death may result.
Q
The system contains a lithium ion battery pack. Replacement and disposal shall be performed only by a qualified service engineer.
Please contact Bayer for assistance.
Compromised Efficacy Hazard - Serious patient injury or death may result.
Q
Do not use this system to deliver any fluid other than intravenous MR contrast media and saline.
Q
Do not retract pistons with connector tube installed.
Procedure Delay Hazard - Serious patient and/or worker injury or death may result.
Q
Turn off any other equipment that could generate high electromagnetic fields or high-level electrostatic discharge.
Personal injury or equipment damage may result when standing on the Power Supply Enclosure.
Q
Do not stand on the power supply enclosure.
Summary of Contents for MEDRAD MRXperion
Page 1: ...Operation Manual...
Page 2: ......
Page 4: ...MEDRAD MRXperion...
Page 10: ...vi MEDRAD MRXperion...
Page 18: ...2 8 MEDRAD MRXperion...
Page 22: ...3 12 MEDRAD MRXperion...
Page 30: ...4 20 MEDRAD MRXperion...
Page 38: ...5 28 MEDRAD MRXperion...
Page 42: ...6 32 MEDRAD MRXperion...
Page 46: ...7 36 MEDRAD MRXperion...
Page 62: ...10 52 MEDRAD MRXperion...
Page 78: ...13 68 MEDRAD MRXperion...
Page 80: ...14 70 MEDRAD MRXperion...
Page 92: ...15 82 MEDRAD MRXperion...
Page 106: ...17 96 MEDRAD MRXperion...
Page 118: ...18 108 MEDRAD MRXperion...
Page 125: ......