10-10
10
Appendices
132387 Rev. D, 2006-06
Power Required
110V AC ± 20%, 45 - 65 Hz
(US only)
220V AC ± 20%, 45 - 65 Hz
(Europe only)
240V AC ± 20%, 45 - 65 Hz (UK
only)
100V - 240V AC ± 20%, 45 - 65 Hz
(with in-line lead)
Battery Powered Operation
Size 6 AA non-rechargeable
alkaline batteries
Dimensions
Depth - 272 mm (10.7 inches)
Width - 171 mm (6.7 inches)
Height - 158 mm (6.2 inches)
Weight
Clinitek Status
®
instrument only
(unpacked, without batteries or
power supply) – 1.66 kg (3.65 lb)
Ambient Operating Temperature
Range
18°C to 30°C (64°F to 86°F)
Ambient Operating Humidity
Range
18% to 80% Relative Humidity
(non-condensing)
Appendix C: Specifications
Optimum Operating Temperature
Range
22°C to 26°C (72°F to 79°F)
Optimum Operating Humidity
Range
35% to 55% Relative Humidity
(non-condensing)
Optimum ranges insure that the
reagent results are optimized for
performance. At temperatures
under 22°C (72°F), urobilinogen
and leukocyte results may be
decreased, and at temperatures
above 26°C (79°F), increased.
Altitude:
2000 m (6562 ft)
Installation Category
: II
Pollution Degree
: 2
Instrument Memory
200 Patient test results
200 Patient details (Patient’s Name
and/or Patient Identification)
Safety Standards
The Clinitek Status analyzer is
classed as a Class A computing
device in accordance with Part 15
of FCC Rules.
Appendix C: Specifications
Summary of Contents for Clinitek Status
Page 1: ...Operator s Manual 132387 Rev D 2006 06 Analyzer...
Page 2: ......
Page 3: ...132387 Rev B 2004 10 Operator s Manual Analyzer...
Page 26: ...2 6 2 Interacting with the Touch Screen 132387 Rev B 2004 10...
Page 28: ...3 2 3 Start Up Wizard 132387 Rev B 2004 10...
Page 80: ...6 4 6 Recall Results 132387 Rev B 2004 10...
Page 90: ...7 10 7 Troubleshooting 132387 Rev B 2004 10...
Page 117: ...10 17 10 Appendices 132387 Rev A 2003 07 Appendix E System Overview Principles...
Page 118: ...10 18 10 Appendices 132387 Rev B 2004 10 Appendix E System Overview Principles...
Page 119: ...10 19 10 Appendices 132387 Rev B 2004 10 Appendix E System Overview Principles...
Page 120: ...10 20 10 Appendices 132387 Rev B 2004 10 Appendix E System Overview Principles...