7-4
7
Troubleshooting
132387 Rev. B, 2004-10
Error Code
Description
Action
E01
Replace the batteries: a) To view instructions
on the display, touch the Error Report selection
area, or b) To use the instructions in this
manual, see page 9-7,
Changing Batteries
.
Low battery power
1. Switch the instrument off by pressing the
on/off button for 2 seconds. 2. Switch the
instrument on again by pressing the on/off
button. 3. Repeat the test.
Loss of test results
E11
1. Make sure that the test table is in place.
Move the test table in or out of the instrument
slightly to reposition the test table. 2. If the
error remains, with the instrument powered
on, unplug the power cord from rear of
instrument and plug back in. Turn instrument
on by pressing the gray power button. 3. If the
error remains with the test table in place,
contact your local Bayer Customer Service
and Support Centre (contact information is
given in Appendix A).
Failure of test table
E23
Replace the batteries: a) To view instructions
on the display, touch the Error Report
selection area, or b) To use the instructions in
this manual, see page 1-2,
Installing Batteries
and/or page 9-7,
Changing Batteries
. If the
battery level becomes too low to power the
instrument, Error Code E01 will be displayed.
Low battery power
E24
Replace the printer paper a) See instructions
on the inside of the printer paper compartment
cover, or b) To view instructions on the display,
touch the Error Report selection area, or c) To
use the instructions in this manual, see page
1-4,
Loading the Printer Paper or Label Roll.
No printer paper
E25,
E64 or
E65
Clean the calibration strip. If the error remains
after cleaning, contact your local Bayer
Customer Service and Support Center
(contact information is given in Appendix A).
Failure of automatic
calibration
E27
1. Switch the instrument off by pressing the
on/off button for 2 seconds. 2. Switch the
instrument on again by pressing the on/off
button.
Set Up failure
Lists of Errors and Advisory Messages
List of Errors and Advisory Messages
E10 or E48
Clinitek Status Analyzer:
Summary of Contents for Clinitek Status
Page 1: ...Operator s Manual 132387 Rev D 2006 06 Analyzer...
Page 2: ......
Page 3: ...132387 Rev B 2004 10 Operator s Manual Analyzer...
Page 26: ...2 6 2 Interacting with the Touch Screen 132387 Rev B 2004 10...
Page 28: ...3 2 3 Start Up Wizard 132387 Rev B 2004 10...
Page 80: ...6 4 6 Recall Results 132387 Rev B 2004 10...
Page 90: ...7 10 7 Troubleshooting 132387 Rev B 2004 10...
Page 117: ...10 17 10 Appendices 132387 Rev A 2003 07 Appendix E System Overview Principles...
Page 118: ...10 18 10 Appendices 132387 Rev B 2004 10 Appendix E System Overview Principles...
Page 119: ...10 19 10 Appendices 132387 Rev B 2004 10 Appendix E System Overview Principles...
Page 120: ...10 20 10 Appendices 132387 Rev B 2004 10 Appendix E System Overview Principles...