9-5
9
Cleaning and Maintenance
132387 Rev. D, 2006-06
Care should be taken not to
scratch the white calibration bar.
Solvents of any kind must not be
used to clean the bar.
Cleaning
4
If the white calibration bar is
clean and unmarked, replace
the table into the analyzer by
holding the table at the end
opposite the white calibration
bar, with the white calibration
bar facing upwards. Push the
test table firmly but slowly, just
over half way into the analyzer.
5
Replace the test table insert.
6
If the white calibration bar is
dirty or discolored, gently wipe
and clean it with a new cotton-
tipped stick or lint-free cloth
wetted with distilled water.
7
Allow the calibration bar to air
dry and then inspect the surface
for dust, foreign material,
scratches or scuffs. If the
calibration bar cannot be
cleaned or is still marked, obtain
a new test table.
8
Reinsert the test table as
described earlier in point 4.
Do not push the test table fully
into the analyzer as the test table
may become jammed and prevent
the use of the analyzer.
Summary of Contents for Clinitek Status
Page 1: ...Operator s Manual 132387 Rev D 2006 06 Analyzer...
Page 2: ......
Page 3: ...132387 Rev B 2004 10 Operator s Manual Analyzer...
Page 26: ...2 6 2 Interacting with the Touch Screen 132387 Rev B 2004 10...
Page 28: ...3 2 3 Start Up Wizard 132387 Rev B 2004 10...
Page 80: ...6 4 6 Recall Results 132387 Rev B 2004 10...
Page 90: ...7 10 7 Troubleshooting 132387 Rev B 2004 10...
Page 117: ...10 17 10 Appendices 132387 Rev A 2003 07 Appendix E System Overview Principles...
Page 118: ...10 18 10 Appendices 132387 Rev B 2004 10 Appendix E System Overview Principles...
Page 119: ...10 19 10 Appendices 132387 Rev B 2004 10 Appendix E System Overview Principles...
Page 120: ...10 20 10 Appendices 132387 Rev B 2004 10 Appendix E System Overview Principles...