TROUBLESHOOTING STAFF CONTROL: STAFF CONTROL LIGHTS BLINK; 

NOTHING ON STAFF CONTROL WORKS, BUT HAND CONTROL PENDANT WORKS

!

 

ALWAYS UNPLUG THE POWER CORD 

BEFORE PERFORMING MAINTENANCE ON 

THE BED.

1.  This could mean the Staff Control is not 

getting enough power from the Control Box.

2.  Unplug the power cord from the wall outlet. 

Unplug the Staff Control Cable from the 
T-Cable at the foot end of the bed.

3.  If you have a spare Staff Control Assembly, 

plug the cable into the foot end T-cable and 
test.

a) 

STAFF CONTROL WORKS AND 
LIGHTS NO LONGER BLINK:

 Replace 

the Staff Control Assembly.

b) 

STAFF CONTROL DOESN’T WORK 
AND LIGHTS STILL BLINK:

 Continue 

Troubleshooting below.

 

Note: Staff Control with Advanced Positioning 

functions is shown above. Operation of Staff 

Control with Standard Positioning functions is 

identical.

TROUBLESHOOTING STAFF CONTROL: NOTHING ON STAFF CONTROL WORKS, 

BUT HAND CONTROL PENDANT WORKS

1.  Check the Staff Control connections at the foot end. 

Is the Staff Control cable plugged securely into the 
T-cable? Also check if the other end of the T-cable is 
plugged securely into the Control Box (or Underbed 
Light). If the 

STAFF CONTROL STILL DOESN’T 

WORK:

 go to step 2.

2.  Unplug the power cord from the wall outlet. Remove 

the Cable Cover (Shroud) on the inside of the 
footboard and locate the single terminal end that 
plugs directly into the back of the Staff Control Panel. 
Ensure it is properly seated in the panel. Plug in the 
power cord and test the Staff Control. If the 

STAFF 

CONTROL STILL DOESN’T WORK:

 Replace the 

Staff Control Assembly.

GF Health Products, Inc. - www.grahamfield.com 

Matrix Series Instructions for Use 999-0908-190A

MATRIX SERIES

A Graham-Field

®

 Family Brand

32

Summary of Contents for Matrix 4000 Series

Page 1: ...use GF Health Products Inc www grahamfield com Matrix Series Instructions for Use 999 0908 190A APR 2017 INSTRUCTIONS FOR USE FDA Recognized Standard ANSI AAMI STD ES60601 1 Health Canada Recognized...

Page 2: ...SE TRENDELENBURG POSITION 27 MATRIX 5000 SERIES BED OPERATION STATIONARY LOW ROLL POSITIONING 28 TROUBLESHOOTING 31 LIMITED WARRANTY 34 These Instructions for Use cover the Matrix Series with Standard...

Page 3: ...bed This especially applies when repositioning or transferring a patient in or out of the bed Increased risk may occur when the patient s size and or weight are inappropriate for the bed s dimensions...

Page 4: ...MAGNETIC COMPATIBILITY EMC INFORMATION WARNING Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information prov...

Page 5: ...e guidelines from a manufacturing aspect However entrapment issues can often arise when a healthcare provider facility has not correctly assembled the components on a bed It is essential that the prov...

Page 6: ...edal Locking Mechanism Check the pedal locking mechanism to ensure it engages and disengages properly X Check the casters and stationary foot pads on both the head end and foot end for any damage wear...

Page 7: ...ass 1 Type B Power Electrical Cord 18 AWG 3 Conductor Type SJT Mode of Operations 10 Max Duty Cycle 2 minutes on 18 minutes off Battery Pack and Charger can be purchased separately as accessories TYPI...

Page 8: ...DMINISTRATION D OXYGENE DE TYPE NASAL OU MASQUE OU AVEC UNE TENTE RECOUVRANT SEULEMENT LA MOTTIE AVENT TETE DU LIT LES COTES DE LAS TENTE OXYGENE NE DOIVENT PAS SE PROLONGER PLUS DAS QUE LA SOMMIER DU...

Page 9: ...MATRIX 4000 SERIES SHOWN DISCARD 1 Foam Block 4 2 Honeycomb Block 8 3 Shipping Board 1 4 Cable Tie 3 Cut and Discard KEEP 5 Panel Mount Kit 6 Mattress Retainer 7 Pendant Holster 8 Wireform Wallsaver N...

Page 10: ...loosely hand tight until after brackets are fastened to the bed frame 5 At the head deck end slide the brackets into the main frame rail hollow ends 6 FOR 80 BEDS Slide the brackets in until the brack...

Page 11: ...m 6 1 a 4 Length Extension Kit is available Please contact Customer Service for more information 5 At the foot deck end slide the brackets into the main frame rail hollow ends 6 FOR 80 BEDS Slide the...

Page 12: ...after brackets are fastened to the bed frame 5 At the head deck end slide the brackets into the main frame rail hollow ends 6 FOR 80 BEDS Slide the brackets in until the first bracket hole aligns wit...

Page 13: ...cm 6 1 a 4 Length Extension Kit is available Please contact Customer Service for more information 5 At the foot deck end slide the brackets into the main frame rail hollow ends 6 FOR 80 BEDS Slide th...

Page 14: ...Carefully squeeze the Mattress Retainer ends inward toward the center of the retainer and insert the ends into the proper holes Lay the Mattress Retainer down so that the elbows rest on the decks and...

Page 15: ...ut of the caster base holes WALLSAVER POSITIONS There are two positions for the Matrix series wallsaver Holes nearest the center of the bed are for both 76 and 80 beds and the holes closest to the hea...

Page 16: ...ds toward the center of the wallsaver until the end tabs slide out of the caster base slots WALLSAVER POSITIONS 1 The Matrix 5000 Series wallsaver features two standard positions for easy bed mattress...

Page 17: ...S REFER TO DETAIL A AT RIGHT a Insert the T Cable end extending out the foot end with phone jack into the round plug making sure the phone jack is seated correctly inside the female plug arrow to arro...

Page 18: ...attached with bolts or quick release pins Wide Deck Kit for non Slide W I D E beds Optional wide kit adjusts the bed sleep deck to 39 or 42 width for greater resident comfort Head and foot boards and...

Page 19: ...k kit For APS9000 and APS7000 For Zenith9000 7000 and 5000 Embedded Staff Control Angled for better visibility Two stage lockouts to prevent pendant use or to prevent use of individual or multiple bed...

Page 20: ...apter Kit Trapeze ZA90100 ZA78100 Long Trendelenburg WallSaver Side WallSaver Bumper set of 2 ZA90070 ZA86200 35 ZA89200 42 GF Health Products Inc www grahamfield com Matrix Series Instructions for Us...

Page 21: ...used with Staff Control and or APS Hand Control Pendant ZA90020 ZA90010 Advanced Positioning Hand Control with Trendelenburg Reverse Trendelenburg Advanced Positioning Hand Control with Chair Position...

Page 22: ...g Staff Control ZL90050 Board Mounted IV Socket Kit Edema Foot Ratchet Kit ZA19000 ZA90030 GF Health Products Inc www grahamfield com Matrix Series Instructions for Use 999 0908 190A MATRIX SERIES A G...

Page 23: ...Bed Transport Dolly set ZA89900 unassembled ZA89900 assembled with bed GF Health Products Inc www grahamfield com Matrix Series Instructions for Use 999 0908 190A 23 MATRIX SERIES A Graham Field Fami...

Page 24: ...H STANDARD POSITIONING HAND CONTROL PENDANT 2 6 4 1 5 3 2 6 8 1 5 7 4 3 2 6 8 1 5 7 4 10 3 9 11 11 NOTE The MATRIX Series Hand Control Pendant can be plugged into the T cable plugs on either side of t...

Page 25: ...tions is shown above Staff Control with Standard Positioning functions is shown at left all standard functions operate the same as those defined above GF Health Products Inc www grahamfield com Matrix...

Page 26: ...hair icon button 2 To release the chair feature and return the bed to horizontal position press Flat Bar icon 4 The Chair feature can also be operated using the Hand Control Pendant HEAD END FOOT END...

Page 27: ...aneously press Key 1 and HI LO Locked button 2 2 To move the bed to Trendelenburg position foot end up press button 3 3 To move the bed to Reverse Trendelenburg position head end up press button 4 Mat...

Page 28: ...tionary non moving position or lowering it into a low roll position This is accomplished by lowering the bed until the stationary foot pads rise off the floor and the rolling casters contact the floor...

Page 29: ...Actuator Cable Actuator Cable HI LO Actuator Head End HI LO Actuator Foot End Head Actuator Control Cable Power Cord Connector CONTROL BOX Optional Battery Cable Connector HEAD END FOOT END GF Health...

Page 30: ...ot Actuator Actuator Cable Power Cord Hand Control Pendant Control Box Actuator Cable Actuator Cable Actuator Cable HI LO Actuator Head End HI LO Actuator Foot End Head Actuator Control Cable GF Healt...

Page 31: ...orange lock icon illuminates then that function is locked out Press Key and Function e g Head buttons simultaneously until orange lock icon de illuminates Test function If PENDANT STILL DOESN T WORK...

Page 32: ...of Staff Control with Standard Positioning functions is identical TROUBLESHOOTING STAFF CONTROL NOTHING ON STAFF CONTROL WORKS BUT HAND CONTROL PENDANT WORKS 1 Check the Staff Control connections at...

Page 33: ...Control box Unplug power cord from wall At the control box switch the HI LO motor cable for the potentially bad motor with the cable for the head motor in port 1 of the Control box Re plug bed power c...

Page 34: ...shall expire on the earlier of warranty period or the expiration date OBTAINING WARRANTY SERVICE A GF Customer Service Representative must authorize warranty service Please contact the GF Customer Ser...

Page 35: ...nd supersede all prior negotiations agreements and understandings with respect thereto The recipient of this document hereby acknowledges and represents that it has not relied on any representation as...

Page 36: ...d on our website www grahamfield com 2017 GF Health Products Inc All Rights Reserved Basic American Medical Products Graham Field and Matrix are trademarks of GF Health Products Inc GF Health Products...

Reviews: