The noise on 
the output 
is much 
higher than 
the specified 
noise at low 
frequency

Ambient 
environment is 
noisy

A noisy environment will be picked up by the sensor. 
Noise tests should be carried out in a shielded 
environment.

Sensor 
breakthrough 
not filtered 
properly

The sensor has a high frequency noise component 
referred to as breakthrough in the datasheet. This 
signal at the excitation frequency of the sensor, if 
unfiltered can be aliased creating an apparent low 
frequency noise. Check that a suitable low pass filter 
is used or that the signal is sampled at a suitable 
frequency (at least twice the breakthrough frequency).

8. Care and Maintenance

Surface dirt contamination on the Mag648 and Mag649™ probe should be removed using 
a mild detergent solution only. Electronics should be cleaned with an antistatic cloth only.

 

ESD protection should be used when handling the electronics board, to prevent irreparable 
damage.

The Mag648 and Mag649™ electronics board should be treated subject to ESD precautions.

Note: 

Store only within the temperature range specified in the product brochure.

Note: Do not expose to strong magnetic fields while being stored as this can magnetise 
the sensor and affect its offset performance.

8.1. Calibration

Return the Mag648 or Mag649™ to Bartington Instruments for calibration at the recommended 
intervals. Refer to the Calibration Certificate for further details.

9. End of Life Disposal

 This product should not be disposed of in domestic or municipal waste. For information 
about disposing of your sensor safely, check local regulations for disposal of electrical / 
electronic products.

Summary of Contents for Mag648

Page 1: ...www bartington com INNOVATION IN MAGNETICS OM2298 4 Operation Manual for Mag648 and Mag649 Low Power Three Axis Magnetic Field Sensors...

Page 2: ...l Description 4 4 1 Vector Measurements and Conventions 4 5 Connections 5 5 1 Cable Recommendations 5 5 2 Connection to Power Supply Acquisition Unit 5 5 4 Mounting Recommendations 6 6 Using the Mag64...

Page 3: ...s a situation in which bodily injury or damage to your instrument or both could result if the caution is ignored Indicates a situation in which ESD protection should be used Identifies items that must...

Page 4: ...48 High stability circuitry ensures that a minimum of ten years service should be expected 4 General Description Key 1 Connector cable 2 Sensing Elements 3 Mounting Holes 4 1 Vector Measurements and C...

Page 5: ...ll cables should be mounted securely in place When designing their own cable the following recommendation should be followed ensure that the cables are shielded to prevent them picking up electromagne...

Page 6: ...mage to the connector or magnetometer When aligned correctly hand tighten the connector Refer to the product brochure for detailed information Caution For the Mag648U and Mag649U only Allow for recess...

Page 7: ...not connected or not sufficient to power the sensor Check that the power supply meets the requirement provided in DS2298 and that the connection is as per Connection to Power Supply Acquisition Unit B...

Page 8: ...e Mag648 and Mag649 probe should be removed using a mild detergent solution only Electronics should be cleaned with an antistatic cloth only ESD protection should be used when handling the electronics...

Page 9: ...ngland The copyright of this document is the property of Bartington Instruments Ltd Bartington is a registered trademark of Bartington Instruments Limited in the following countries Australia Brazil C...

Reviews: