background image

BARTINGTON INSTRUMENTS

 

Page 5 of 9 

OM2298/4

North 

+X 

Thumb

Down 

+Z 

2nd Finger

East 

+Y 

1st Finger

5.  Connections

5.1. Cable Recommendations

The Mag648 and Mag649™ electronics board provides balanced output lines for analogue signal 
transmission. Suitable shielded cables can be supplied.

Cables are particularly prone to wear and damage if twisted, flexed beyond their design limits, or 
subjected to excessive or repeated movement. All cables should be mounted securely in place.

When designing their own cable, the following recommendation should be followed:

•  ensure that the cables are shielded to prevent them picking up electromagnetic 

interference.

•  the cable shield should be connected to power supply ground at the power supply end 

(either through the connector body, or by connecting the shield to the power ground 
wire).

5.2. Connection to Power Supply/Acquisition Unit

Caution:

 When providing your own power supply, do not exceed the voltage rating, provide 

sufficient current, and ensure correct polarity is respected see DS2298.

Figure 1. The ‘right hand’ rule.

Summary of Contents for Mag648

Page 1: ...www bartington com INNOVATION IN MAGNETICS OM2298 4 Operation Manual for Mag648 and Mag649 Low Power Three Axis Magnetic Field Sensors...

Page 2: ...l Description 4 4 1 Vector Measurements and Conventions 4 5 Connections 5 5 1 Cable Recommendations 5 5 2 Connection to Power Supply Acquisition Unit 5 5 4 Mounting Recommendations 6 6 Using the Mag64...

Page 3: ...s a situation in which bodily injury or damage to your instrument or both could result if the caution is ignored Indicates a situation in which ESD protection should be used Identifies items that must...

Page 4: ...48 High stability circuitry ensures that a minimum of ten years service should be expected 4 General Description Key 1 Connector cable 2 Sensing Elements 3 Mounting Holes 4 1 Vector Measurements and C...

Page 5: ...ll cables should be mounted securely in place When designing their own cable the following recommendation should be followed ensure that the cables are shielded to prevent them picking up electromagne...

Page 6: ...mage to the connector or magnetometer When aligned correctly hand tighten the connector Refer to the product brochure for detailed information Caution For the Mag648U and Mag649U only Allow for recess...

Page 7: ...not connected or not sufficient to power the sensor Check that the power supply meets the requirement provided in DS2298 and that the connection is as per Connection to Power Supply Acquisition Unit B...

Page 8: ...e Mag648 and Mag649 probe should be removed using a mild detergent solution only Electronics should be cleaned with an antistatic cloth only ESD protection should be used when handling the electronics...

Page 9: ...ngland The copyright of this document is the property of Bartington Instruments Ltd Bartington is a registered trademark of Bartington Instruments Limited in the following countries Australia Brazil C...