background image

BARTINGTON INSTRUMENTS

 

Page 7 of 9 

OM2298/4

6.2. Electromagnetic Compatibility

Note: 

The Mag648 and Mag649™ sensors are not shielded for immunity from, or emission 

of, electromagnetic fields. 

7. Troubleshooting

The sensor is unlikely to suffer any defects in normal use: no internal components are 
serviceable. The most likely causes of failure, and their solutions, are detailed in the following 
table. 

In the event of any apparent malfunction beyond those described in the table below, please email 

[email protected]

, or telephone the Bartington Instruments service team on +44 (0)1993 

706565. 

Problem

Cause

Solution

There are 
no analogue 
outputs 
present in X, Y 
or Z

Sensor head not 
connected

Ensure that the sensor head and electronics are 
connected as per 

Probe and Electronics wiring

.

Power supply 
not connected or 
not sufficient to 
power the sensor

Check that the power supply meets the requirement 
provided in DS2298 and that the connection is as per 

Connection to Power Supply/Acquisition Unit

.

Broken coil

If only one or two axes is (are) not responding, one of the 
fluxgate coil could be damaged. Please return the unit to 
us for assessment.

Component 
damage

If only one or two axes is (are) not responding, one 
component could be damaged. Please return the unit to 
us for assessment.

There is an 
abnormal 
current draw 
on the sensor

Faulty 
component

A faulty component can lead to an abnormal current 
draw before it fails. Please return the unit to us for 
assessment.

The 
amplitude of 
the Earth’s 
field is 
abnormal

Erroneous 
scaling factor

If the scaling factor applied to convert volts into 
field amplitude is erroneous, the field reading will 
appear either systematically high or low depending 
on the error on the scaling factor. Please check the 
conversion rate used in the acquisition software.

Summary of Contents for Mag648

Page 1: ...www bartington com INNOVATION IN MAGNETICS OM2298 4 Operation Manual for Mag648 and Mag649 Low Power Three Axis Magnetic Field Sensors...

Page 2: ...l Description 4 4 1 Vector Measurements and Conventions 4 5 Connections 5 5 1 Cable Recommendations 5 5 2 Connection to Power Supply Acquisition Unit 5 5 4 Mounting Recommendations 6 6 Using the Mag64...

Page 3: ...s a situation in which bodily injury or damage to your instrument or both could result if the caution is ignored Indicates a situation in which ESD protection should be used Identifies items that must...

Page 4: ...48 High stability circuitry ensures that a minimum of ten years service should be expected 4 General Description Key 1 Connector cable 2 Sensing Elements 3 Mounting Holes 4 1 Vector Measurements and C...

Page 5: ...ll cables should be mounted securely in place When designing their own cable the following recommendation should be followed ensure that the cables are shielded to prevent them picking up electromagne...

Page 6: ...mage to the connector or magnetometer When aligned correctly hand tighten the connector Refer to the product brochure for detailed information Caution For the Mag648U and Mag649U only Allow for recess...

Page 7: ...not connected or not sufficient to power the sensor Check that the power supply meets the requirement provided in DS2298 and that the connection is as per Connection to Power Supply Acquisition Unit B...

Page 8: ...e Mag648 and Mag649 probe should be removed using a mild detergent solution only Electronics should be cleaned with an antistatic cloth only ESD protection should be used when handling the electronics...

Page 9: ...ngland The copyright of this document is the property of Bartington Instruments Ltd Bartington is a registered trademark of Bartington Instruments Limited in the following countries Australia Brazil C...

Reviews: