BARTINGTON INSTRUMENTS
Page 14 of 21 OM3224/4
Device Input gain
Signal
PXI function
Pin
Device Under Test Input Gain
P0.10
49
0
Device Under Test Input Gain = 100
1
Device Under Test Input Gain = 1
Input Filter
Signal
PXI function
Pin
Filter Enable
P0.29
40
Filter Select 0
P0.30
1
Filter select 1
P0.31
39
Select Inputs
Nominal frequency at –3dB
point of Selected filter
0
1
EN
0
0
0
10Hz
1
0
0
100Hz
0
1
0
1KHz
1
1
0
10KHz
X
X
0
No Filter
Power Supply Monitor
Signal
PXI function
Pin
PSU Monitor Enable
P0.21
6
PSU Monitor Select 0
P0.16
11
PSU Monitor Select 1
P0.17
10
PSU Monitor Select 2
P0.18
43
0
1
2
EN
X
X
X
0
Output Disabled (voltage uncertain)
0
0
1
1
¬Device under test negative supply
voltage
1
0
1
1
Device under test negative supply
current
0
1
1
1
¬Device under test positive supply
Voltage
1
1
1
1
¬Device under test positive supply
current
X
X
0
1
Invalid input (voltage uncertain)