LABDMM
2
USER GUIDE
MO.LABDMM2.560.ENG.R1
Pag. 22
30. CALIBRATION OF THE ATMOSPHERIC PRESSURE
This calibration can only be performed only for ABSOLUTE
manometers.
+
Press the keys simultaneously for a few seconds
SET
and
PEAK.
P0000
Set the password
0022
(
) and confirm with
SET.
1.0000
In this phase it is possible to enter the value of the atmospheric pressure.
The last stored calibration atmopheric pressure value (in bar) will be
displayed.
It is recommended to refer to a reliable detection, in order not to invalidate
the reading with an incorrect value.
Press
SET
to exit the function and return to the pressure measurement
WARNING
The calibration procedures are reported in this manual only as documentation and must be performed only by
authorized calibration centers and in case of actual need.
The incorrect alteration of these parameters will make the following measurements unreliable.
BART
declines all responsibility for measurement errors or malfunctions that may result from
incorrectly performed adjustments, which also invalidate the
ACCREDIA
certification of the pressure gauge.