19. PARAMETER MENU
To enter the Parameter Menu keep the
SET
key pressed for about 3 seconds, until the writing
Unit
appears.
Unit
PRESSURE UNIT SELECTION
In this step it is possible to change the pressure unit of measurement.
Select the unit using the
and
keys.
Confirm with
SET
to go to the next parameter
.
FL XX
DIGITAL FILTER
In this step it is possible to vary the
DIGITAL FILTER
effect by pressing the
and
keys.
Increasing the
XX
value increases the filter effect, allowing the user to detect the
mean value of unstable or pulsating measurements.
The selectable values are from 0 to 5.
Confirm with
SET
to go to the next parameter.
r XX
RESOLUTION
In this step it is possible to program the
RESOLUTION
with which the instrument
displays the output.
Selectable values 1, 2, 5 and 10.
Modify the parameter using the
and
keys.
Confirm with
SET
to go to the next parameter.
LOOPX
The
LOOP
function allows you to view both pressure and temperature.
By setting X=1, the LOOP function is activated.
On the measurement page every 30 seconds the pressure and temperature will
be displayed alternately.
By setting X=0, the function is deactivated
Modify the parameter using the
and
keys.
Confirm with
SET
to go to the next parameter.
oFFXX
AUTO POWER OFF TIME
Defines the number of minutes (from 1 to 90) before the automatic switch-off in
case of constant pressure.
The power off time starts working if the pressure gauge does not detect pressure
variations greater than 10% of the full scale.
Modify the parameter using the
and
keys.
Confirm with
SET
to go to the next parameter.
cont X
CONTINUOUS SERIAL TRANSMISSION MODE : ON / OFF
cont0
: serial pressure transmission is obtained on request
cont1
: the pressure value is transmitted continuously every 100ms.
Modify the parameter using the
and
keys.
Confirm with
SET
to go to the next parameter.
Lt XX
BACKLIGHT ACTIVATION TIME
This parameter defines the switching on time of the backlight activated when the
SET
key is pressed for a short time.
The time can be set between 1s and 99s. By setting 0, the backlight is turned off.
Activating the backlight will result in higher battery consumption, so it is better
to deactivate the function when it is not used.
ZEroX
Enable the ZERO function (only for Absolute manometers)
Set
ZEro0 to
di
s
able
the zero function (default)
Set
Zero1 to en
able
the zero function
Summary of Contents for LABDMM2
Page 5: ......
Page 6: ......
Page 7: ......
Page 18: ......
Page 27: ...LABDMM2 Manuale Operativo MO LABDMM2 560 R0 Pag 26 33 Dimensioni mm CONTENITORE DA INCASSO...
Page 28: ...MO LABDMM2 560 ENG R1 PROFESSIONAL DIGITAL MANOMETER for PRESSURE and TEMPERATURE measurement...
Page 57: ...LABDMM2 USER GUIDE MO LABDMM2 560 ENG R1 Pag 29 34 Dimensions mm BUILT IN VERSION...