19
High Temperature Cut-off Switch
Your MEGA-PURE One Liter Water Still is protected against
overheating by a thermal switch located in the control box at
the right hand side of the boiler. Should the boiler overheat,
the switch will open causing the heater and water supply to
shut off. When the boiler cools (5-15 minutes), the switch will
reset automatically, but the still will have to be restarted by
the operator. When unit has cooled, press the “WATER”
switch and then the “OPERATE” switch to restore normal
operation. Check boiler occasionally for proper water level.
O
PERATION
Note
If you are feeding an ACS with your
still, note that when the ACS is full, it
will signal the still to shut down.
During this shut-down, the heater
and the water supply will both shut
off and the front panel lights on the
still will be extinguished. This normal
operation can mimic the effects of
an overheating boiler, but does not
represent any problem with the still.